Literature DB >> 23945151

Towards true unipolar ECG recording without the Wilson central terminal (preliminary results).

G D Gargiulo1, A L McEwan, P Bifulco, M Cesarelli, C Jin, J Tapson, A Thiagalingam, A van Schaik.   

Abstract

We present an innovative bio-potential front-end capable of recording true unipolar ECG leads for the first time without making use of the Wilson central terminal. In addition to the convenience in applications such as continuous monitoring and rapid diagnosis, the information in unipolar recordings may yield unique diagnostic information as it avoids the need to essentially subtract data or make use of the averaging effect imposed from the Wilson central terminal. The system also allows direct, real-time software calculation of signals corresponding to standard ECG leads which achieve correlations in excess of 92% with a gold standard ECG during a parallel in vivo recording. In addition, the implemented circuit is wideband (0.05-1000 Hz), compatible with standard (Ag/AgCl) bio-potential electrodes, and dry (paste-less) textile electrodes. The circuit is also low power, requiring less than 50 mW (when powered at 12 V) per standard ECG lead (two channels required). It is therefore well suited for wearable, long-term applications.

Mesh:

Year:  2013        PMID: 23945151     DOI: 10.1088/0967-3334/34/9/991

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  5 in total

Review 1.  Individual identification via electrocardiogram analysis.

Authors:  Antonio Fratini; Mario Sansone; Paolo Bifulco; Mario Cesarelli
Journal:  Biomed Eng Online       Date:  2015-08-14       Impact factor: 2.819

2.  WCTECGdb: A 12-Lead Electrocardiography Dataset Recorded Simultaneously with Raw Exploring Electrodes' Potential Directly Referred to the Right Leg.

Authors:  Hossein Moeinzadeh; Joseph Assad; Paolo Bifulco; Mario Cesarelli; Aiden O'Loughlin; Jonathan C Tapson; Ibrahim M Shugman; Aravinda Thiagalingam; Gaetano D Gargiulo
Journal:  Sensors (Basel)       Date:  2020-06-08       Impact factor: 3.576

3.  Real-Time Ventricular Volume Measured Using the Intracardiac Electromyogram.

Authors:  Marianne Schmid Daners; Sophie Hall; Simon Sündermann; Nikola Cesarovic; Mareike Kron; Volkmar Falk; Christoph Starck; Mirko Meboldt; Seraina A Dual
Journal:  ASAIO J       Date:  2021-12-01       Impact factor: 3.826

4.  True unipolar ECG machine for Wilson Central Terminal measurements.

Authors:  Gaetano D Gargiulo
Journal:  Biomed Res Int       Date:  2015-10-01       Impact factor: 3.411

5.  On the Einthoven Triangle: A Critical Analysis of the Single Rotating Dipole Hypothesis.

Authors:  Gaetano D Gargiulo; Paolo Bifulco; Mario Cesarelli; Alistair L McEwan; Hossein Moeinzadeh; Aiden O'Loughlin; Ibrahim M Shugman; Jonathan C Tapson; Aravinda Thiagalingam
Journal:  Sensors (Basel)       Date:  2018-07-20       Impact factor: 3.576