Literature DB >> 23941656

First clinical experience with Celt ACD(®) : a femoral arterial puncture closure device.

Aftab Jan1, Peter Crean, Lutz Bullesfeld, James Coleman, Eberhard Grube, Niall Mulvihill.   

Abstract

INTRODUCTION: This prospective nonrandomized study compared the safety and efficacy of a novel arterial closure device (ACD) in common femoral artery procedures to that of the FDA submitted historical manual pressure control group, who underwent either a diagnostic angiogram (DA) or a percutaneous coronary intervention (PCI) procedure. METHODS AND
RESULTS: A total of 55 patients were enrolled in this study of the novel ACD. Of the 55 patients, 39 were enrolled in the DA group and 16 were enrolled in the PCI group. Six patients were excluded. A device was deployed in 49 patients. Time to hemostasis (TTH), time to ambulation (TTA), device function, and device-related vascular complications were measured. In the device group, the TTH for the combined DA and PCI patients was 32 seconds (0.54 ± 0.93 minutes), significantly lower when compared with 16.0 ± 12.2 minutes (P<0.0001) for the control group. Overall major vascular complication rate did not differ significantly, device group (1/49) and the historical control group (1/217). TTA in the combined PCI and DA device group was 226.4 ± 231.9 at the German site (site ambulation policy). In the Irish site, the average TTA in the PCI group was 187 minutes (n=8) and 85 minutes (n=14) in the DA group.
CONCLUSION: The Celt ACD® device is safe, effective, and significantly decreases the TTH compared to manual pressure and has a low vascular complications rate. The device may be effective in early ambulation and discharge of patients postcoronary intervention procedures.
© 2013, Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2013        PMID: 23941656     DOI: 10.1111/joic.12046

Source DB:  PubMed          Journal:  J Interv Cardiol        ISSN: 0896-4327            Impact factor:   2.279


  4 in total

1.  Effects of position change on lumbar pain and discomfort of Korean patients after invasive percutaneous coronary intervention: a RCT study.

Authors:  Nam Hyun Cha; Sohyune Sok
Journal:  J Phys Ther Sci       Date:  2016-10-28

2.  Successful surgical retrieval of Celt ACD® vascular closure device embolised in the tibioperoneal trunk.

Authors:  Qusai Aljarrah; Ma'moon Al-Omari; Kawthar Qader; Jozef Oweis; Ahmad Althaher
Journal:  CVIR Endovasc       Date:  2018-08-07

3.  Maldeployment of Celt ACD vascular closure device.

Authors:  Benjamin Hart; Sachinder Singh Hans
Journal:  J Vasc Surg Cases Innov Tech       Date:  2021-12-09

4.  Endovascular Management of Superficial Femoral Artery Occlusion Secondary to Embolization of Celt ACD® Vascular Closure Device.

Authors:  Jehad Fataftah; Raed Tayyem; Bashar Ghosheh; Farah Tayyem; Mamoon Al-Omari
Journal:  Am J Case Rep       Date:  2020-08-27
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.