Evaluation of the automated hematology analyzer ADVIA® 120 for cerebrospinal fluid analysis and usage of unique hemolysis reagent.

H Tanada1, T Ikemoto1, R Masutani1, H Tanaka1, T Takubo2.   

Abstract

INTRODUCTION: In this study, we evaluated the performance of the ADVIA 120 hematology system for cerebrospinal fluid (CSF) assay.
METHODS: Cell counts and leukocyte differentials in CSF were examined with the ADVIA 120 hematology system, while simultaneously confirming an effective hemolysis agent for automated CSF cell counts.
RESULTS: The detection limits of both white blood cell (WBC) counts and red blood cell (RBC) counts on the measurement of CSF cell counts by the ADVIA 120 hematology system were superior at 2 cells/μL (10(-6) L). The WBC count was linear up to 9.850 cells/μL, and the RBC count was linear up to approximately 20 000 cells/μL. The intrarun reproducibility indicated good precision. The leukocyte differential of CSF cells, performed by the ADVIA120 hematology system, showed good correlation with the microscopic procedure. The VersaLyse hemolysis solution efficiently lysed the samples without interfering with cell counts and leukocyte differential, even in a sample that included approximately 50 000/μL RBC.
CONCLUSION: These data show the ADVIA 120 hematology system correctly measured the WBC count and leukocyte differential in CSF. The VersaLyse hemolysis solution is considered to be optimal for hemolysis treatment of CSF when measuring cell counts and differentials by the ADVIA 120 hematology system.
© 2013 John Wiley & Sons Ltd.

Entities:  

Keywords:  ADVIA 120; Cerebrospinal fluid; VersaLyse; hemolysis

Mesh:

Year:  2013        PMID: 23937551     DOI: 10.1111/ijlh.12130

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  2 in total

1.  The World Health Organization's Essential Diagnostics List: Diagnostics for neurologic disorders.

Authors:  Greer Waldrop; Teddy G Goetz; Omar K Siddiqi; Igor J Koralnik; Hiral Shah; Kiran T Thakur
Journal:  Neurology       Date:  2019-10-08       Impact factor: 9.910

2.  Automated Analysis of Cerebrospinal Fluid Cells Using Commercially Available Blood Cell Analysis Devices-A Critical Appraisal.

Authors:  Manfred Wick; Catharina C Gross; Hayrettin Tumani; Brigitte Wildemann; Martin Stangel
Journal:  Cells       Date:  2021-05-18       Impact factor: 6.600

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.