BACKGROUND: We compared a novel assay for free light chain (FLC) quantitation based on monoclonal antibodies (N-Latex, Siemens, Germany) to the established polyclonal antibody-based assay (Freelite™, The Binding Site, UK) in AL amyloidosis. METHODS: Sixty-two diagnostic samples were analysed on a BNII nephelometer, 32 of which also had a post-treatment sample. RESULTS: In the diagnostic samples: for AL of κ type, the median involved FLC (iFLC) was significantly lower by the N-Latex assay (289 vs. 667 mg/L, p=0.0002) whereas for λ AL the values were similar (148 vs. 161 mg/L, p=0.84). Measurable disease, defined as a difference between involved and uninvolved FLC (dFLC) >50 mg/L was present in 82% by the N-Latex assay compared to 89% by the Freelite™ assay. For diagnostic sensitivity, the FLC ratio was normal in 21% (95% CI 12%-33%) and 15% (95% CI 7%-26%) of patients by the N-Latex and Freelite™ assays, respectively. The combination of serum and urine immunofixation electrophoresis with either FLC assay allowed identification of the amyloidogenic clone in 98% producing comparable sensitivity. For the monitoring samples the median reduction in dFLC was 68% for the N-Latex assay and 77% for the Freelite™ assay (p=0.04). This led to some differences in assigning response categories. Partial response as assigned by both assays predicted overall survival (N-Latex p=0.0015, Freelite™ p=0.022). CONCLUSIONS: There are differences between FLC as measured by the N-Latex and Freelite™ assays, but overall the two assays have similar diagnostic sensitivity. Disease response calculated by both assays predicts survival but more clinical validation is required.
BACKGROUND: We compared a novel assay for free light chain (FLC) quantitation based on monoclonal antibodies (N-Latex, Siemens, Germany) to the established polyclonal antibody-based assay (Freelite™, The Binding Site, UK) in AL amyloidosis. METHODS: Sixty-two diagnostic samples were analysed on a BNII nephelometer, 32 of which also had a post-treatment sample. RESULTS: In the diagnostic samples: for AL of κ type, the median involved FLC (iFLC) was significantly lower by the N-Latex assay (289 vs. 667 mg/L, p=0.0002) whereas for λ AL the values were similar (148 vs. 161 mg/L, p=0.84). Measurable disease, defined as a difference between involved and uninvolved FLC (dFLC) >50 mg/L was present in 82% by the N-Latex assay compared to 89% by the Freelite™ assay. For diagnostic sensitivity, the FLC ratio was normal in 21% (95% CI 12%-33%) and 15% (95% CI 7%-26%) of patients by the N-Latex and Freelite™ assays, respectively. The combination of serum and urine immunofixation electrophoresis with either FLC assay allowed identification of the amyloidogenic clone in 98% producing comparable sensitivity. For the monitoring samples the median reduction in dFLC was 68% for the N-Latex assay and 77% for the Freelite™ assay (p=0.04). This led to some differences in assigning response categories. Partial response as assigned by both assays predicted overall survival (N-Latex p=0.0015, Freelite™ p=0.022). CONCLUSIONS: There are differences between FLC as measured by the N-Latex and Freelite™ assays, but overall the two assays have similar diagnostic sensitivity. Disease response calculated by both assays predicts survival but more clinical validation is required.
Authors: Donna E Reece; Ute Hegenbart; Vaishali Sanchorawala; Giampaolo Merlini; Giovanni Palladini; Joan Bladé; Jean-Paul Fermand; Hani Hassoun; Leonard Heffner; Vishal Kukreti; Robert A Vescio; Lixia Pei; Christopher Enny; Dixie-Lee Esseltine; Helgi van de Velde; Andrew Cakana; Raymond L Comenzo Journal: Blood Date: 2014-09-08 Impact factor: 22.113