Amy L McPherson1, Glenn M Walker.
Abstract
A microfluidic device using on-chip passive pumping was characterized for use as a particle counter. Flow occurred due to a Young-Laplace pressure gradient between two 1.2 mm diameter inlets and a 4 mm diameter reservoir when 0.5μ L fluid droplets were applied to the inlets using a micropipette. Polystyrene particles (10μm diameter) were enumerated using the resistive pulse technique. Particle counts using passive pumping were within 13% of counts from a device using syringe pumping. All pumping methods produced particle counts that were within 16% of those obtained with a hemocytometer. The effect of intermediate wash steps on particle counts within the passive pumping device was determined. Zero, one, or two wash droplets were loaded after the first of two sample droplets. No statistical difference was detected in the mean particle counts among the loading patterns (p > 0.05). Hydrodynamic focusing using passive pumping was also demonstrated.Entities:
Keywords: Colloid; Enumeration; Microfluidics; Resistive Pulse
Year: 2010 PMID: 23930109 PMCID: PMC3735229 DOI: 10.1007/s10404-010-0609-0
Source DB: PubMed Journal: Microfluid Nanofluidics ISSN: 1613-4982 Impact factor: 2.529