Performance parameters of a liquid filled ionization chamber array.

B Poppe1, T S Stelljes, H K Looe, N Chofor, D Harder, K Willborn.   

Abstract

PURPOSE: In this work, the properties of the two-dimensional liquid filled ionization chamber array Octavius 1000SRS (PTW-Freiburg, Germany) for use in clinical photon-beam dosimetry are investigated.
METHODS: Measurements were carried out at an Elekta Synergy and Siemens Primus accelerator. For measurements of stability, linearity, and saturation effects of the 1000SRS array a Semiflex 31013 ionization chamber (PTW-Freiburg, Germany) was used as a reference. The effective point of measurement was determined by TPR measurements of the array in comparison with a Roos chamber (type 31004, PTW-Freiburg, Germany). The response of the array with varying field size and depth of measurement was evaluated using a Semiflex 31010 ionization chamber as a reference. Output factor measurements were carried out with a Semiflex 31010 ionization chamber, a diode (type 60012, PTW-Freiburg, Germany), and the detector array under investigation. The dose response function for a single detector of the array was determined by measuring 1 cm wide slit-beam dose profiles and comparing them against diode-measured profiles. Theoretical aspects of the low pass properties and of the sampling frequency of the detector array were evaluated. Dose profiles measured with the array and the diode detector were compared, and an intensity modulated radiation therapy (IMRT) field was verified using the Gamma-Index method and the visualization of line dose profiles.
RESULTS: The array showed a short and long term stability better than 0.1% and 0.2%, respectively. Fluctuations in linearity were found to be within ±0.2% for the vendor specified dose range. Saturation effects were found to be similar to those reported in other studies for liquid-filled ionization chambers. The detector's relative response varied with field size and depth of measurement, showing a small energy dependence accounting for maximum signal deviations of ±2.6% from the reference condition for the setup used. The σ-values of the Gaussian dose response function for a single detector of the array were found to be (0.72±0.25) mm at 6 MV and (0.74±0.25) mm at 15 MV and the corresponding low pass cutoff frequencies are 0.22 and 0.21 mm(-1), respectively. For the inner 5×5 cm2 region and the outer 11×11 cm2 region of the array the Nyquist theorem is fulfilled for maximum sampling frequencies of 0.2 and 0.1 mm(-1), respectively. An IMRT field verification with a Gamma-Index analysis yielded a passing rate of 95.2% for a 3 mm∕3% criterion with a TPS calculation as reference.
CONCLUSIONS: This study shows the applicability of the Octavius 1000SRS in modern dosimetry. Output factor and dose profile measurements illustrated the applicability of the array in small field and stereotactic dosimetry. The high spatial resolution ensures adequate measurements of dose profiles in regular and intensity modulated photon-beam fields.

Entities:  

Mesh:

Year:  2013        PMID: 23927343     DOI: 10.1118/1.4816298

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  11 in total

1.  Under-reported dosimetry errors due to interplay effects during VMAT dose delivery in extreme hypofractionated stereotactic radiotherapy.

Authors:  Tobias Gauer; Thilo Sothmann; Oliver Blanck; Cordula Petersen; René Werner
Journal:  Strahlenther Onkol       Date:  2018-02-15       Impact factor: 3.621

2.  Operation and calibration of the novel PTW 1600SRS detector for the verification of single isocenter stereotactic radiosurgery treatments of multiple small brain metastases.

Authors:  Esther Decabooter; Ans Cc Swinnen; Michel C Öllers; Fabian Göpfert; Frank Verhaegen
Journal:  Br J Radiol       Date:  2021-06-11       Impact factor: 3.629

3.  4D dose simulation in volumetric arc therapy: Accuracy and affecting parameters.

Authors:  Thilo Sothmann; Tobias Gauer; René Werner
Journal:  PLoS One       Date:  2017-02-23       Impact factor: 3.240

4.  Assessment of Imprecise Small Photon Beam Modeling by Two Treatment Planning System Algorithms.

Authors:  Hadi Keivan; Daryoush Shahbazi-Gahrouei; Ahmad Shanei; Alireza Amouheidari
Journal:  J Med Signals Sens       Date:  2018 Jan-Mar

5.  Evaluation of semiempirical VMAT dose reconstruction on a patient dataset based on biplanar diode array measurements.

Authors:  Cassandra Stambaugh; Daniel Opp; Stuart Wasserman; Geoffrey Zhang; Vladimir Feygelman
Journal:  J Appl Clin Med Phys       Date:  2014-03-06       Impact factor: 2.102

6.  Verification of high-dose-rate brachytherapy treatment planning dose distribution using liquid-filled ionization chamber array.

Authors:  A B Mohamed Yoosuf; Prakash Jeevanandam; Glenn Whitten; Geraldine Workman; Conor K McGarry
Journal:  J Contemp Brachytherapy       Date:  2018-04-30

Review 7.  Influence of the jaw tracking technique on the dose calculation accuracy of small field VMAT plans.

Authors:  Ans C C Swinnen; Michel C Öllers; Erik Roijen; Sebastiaan M Nijsten; Frank Verhaegen
Journal:  J Appl Clin Med Phys       Date:  2017-01-03       Impact factor: 2.102

8.  A hybrid volumetric dose verification method for single-isocenter multiple-target cranial SRS.

Authors:  Saeed Ahmed; Jeff Kapatoes; Geoffrey Zhang; Eduardo G Moros; Vladimir Feygelman
Journal:  J Appl Clin Med Phys       Date:  2018-08-15       Impact factor: 2.102

9.  Clinical Evaluation of a Two-dimensional Liquid-Filled Ion chamber Detector Array for Verification of High Modulation Small Fields in Radiotherapy.

Authors:  Miljenko Markovic; Ganesh Narayanasamy; Sotirios Stathakis; Panayiotis Mavroidis; Ines-Ana Jurkovic; Daniel Saenz; Niko Papanikolaou
Journal:  J Med Phys       Date:  2019 Apr-Jun

10.  Volumetric-modulated Arc Therapy Lung Stereotactic Body Radiation Therapy Dosimetric Quality Assurance: A Comparison between Radiochromic Film and Chamber Array.

Authors:  Juan Fernando Mata Colodro; Alfredo Serna Berná; Vicente Puchades Puchades; David Ramos Amores; Miguel Alcaraz Baños
Journal:  J Med Phys       Date:  2017 Jul-Sep
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.