Cardiac index assessment: validation of a new non-invasive very low current thoracic bioimpedance device by thermodilution.

Andrea Faini1, Stefano Omboni, Marius Tifrea, Serban Bubenek, Ovidiu Lazar, Gianfranco Parati.   

Abstract

INTRODUCTION: The accuracy of impedance cardiography for cardiac index assessment is matter of debate, with available studies reporting inconsistent results. Our study aimed at evaluating the agreement between measurements of cardiac index provided by a new-generation thoracic electrical bioimpedance device (Hotman System) and an invasive approach based on thermodilution in humans.
METHODS: Cardiac index was assessed simultaneously with thoracic electrical bioimpedance and conventional thermodilution through comparison of five consecutive measurements in 51 cardiac patients, hospitalized in an intensive care unit (mean± SD age: 60 ± 11 years; 68% males). The agreement between cardiac index values measured by both methods was assessed by the Bland-Altman approach, adjusted for repeated measures. The repeatability coefficient and the intraclass correlation coefficient were used to assess reproducibility of replicates.
RESULTS: Average (± SD) cardiac index was 3.05 ± 0.91 l/min/m(2) with Hotman System and 3.14 ± 1.12 l/min/m(2) with thermodilution. The bias of precision was -0.09 ± 0.41. The coefficients of repeatability and intraclass correlation coefficients were high and similar for the two techniques (0.95 l/min/m(2) and 0.91 for Hotman System vs 0.78 l/min/m(2) and 0.90 for thermodilution).
CONCLUSIONS: Cardiac index values yielded by Hotman system compares favorably with that obtained with thermodilution in cardiac patients.

Entities:  

Mesh:

Year:  2013        PMID: 23926884     DOI: 10.3109/08037051.2013.817121

Source DB:  PubMed          Journal:  Blood Press        ISSN: 0803-7051            Impact factor:   2.835


  2 in total

1.  Relationships between retinal arteriole anatomy and aortic geometry and function and peripheral resistance in hypertensives.

Authors:  David Rosenbaum; Nadjia Kachenoura; Edouard Koch; Michel Paques; Philippe Cluzel; Alban Redheuil; Xavier Girerd
Journal:  Hypertens Res       Date:  2016-03-24       Impact factor: 3.872

2.  Accuracy, Precision, and Trending Ability of Electrical Cardiometry Cardiac Index versus Continuous Pulmonary Artery Thermodilution Method: A Prospective, Observational Study.

Authors:  P B W Cox; A M den Ouden; M Theunissen; L J Montenij; A G H Kessels; M D Lancé; W F F A Buhre; M A E Marcus
Journal:  Biomed Res Int       Date:  2017-10-09       Impact factor: 3.411

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.