Effect of mechanical behaviour of the brachial artery on blood pressure measurement during both cuff inflation and cuff deflation.

Dingchang Zheng1, Fan Pan, Alan Murray.   

Abstract

OBJECTIVE: The aim of this study was to investigate the effect of different mechanical behaviour of the brachial artery on blood pressure (BP) measurements during cuff inflation and deflation.
METHODS: BP measurements were taken from each of 40 participants, with three repeat sessions under three randomized cuff deflation/inflation conditions. Cuff pressure was linearly deflated and inflated at a standard rate of 2-3 mmHg/s and also linearly inflated at a fast rate of 5-6 mmHg/s. Manual auscultatory systolic and diastolic BPs, and pulse pressure (SBP, DBP, PP) were measured. Automated BPs were determined from digitally recorded cuff pressures by fitting a polynomial model to the oscillometric pulse amplitudes. The BPs from cuff deflation and inflation were then compared.
RESULTS: Repeatable measurements between sessions and between the sequential order of inflation/deflation conditions (all P > 0.1) indicated stability of arterial mechanical behaviour with repeat measurements. Comparing BPs obtained by standard inflation with those from standard deflation, manual SBP was 2.6 mmHg lower (P < 0.01), manual DBP was 1.5 mmHg higher (P < 0.01), manual PP was 4.2 mmHg lower (P < 0.001), automated DBP was 6.7 mmHg higher (P < 0.001) and automatic PP was 7.5 mmHg lower (P < 0.001). There was no statistically significant difference for any automated BPs between fast and standard cuff inflation. The statistically significant BP differences between inflation and deflation suggest different arterial mechanical behaviour between arterial opening and closing during BP measurement.
CONCLUSION: We have shown that the mechanical behaviour of the brachial artery during BP measurement differs between cuff deflation and cuff inflation.

Entities:  

Mesh:

Year:  2013        PMID: 23924706     DOI: 10.1097/MBP.0b013e3283651d55

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  3 in total

1.  Significantly Reduced Blood Pressure Measurement Variability for Both Normotensive and Hypertensive Subjects: Effect of Polynomial Curve Fitting of Oscillometric Pulses.

Authors:  Fangwei Yang; Fei Chen; Mingping Zhu; Aiqing Chen; Dingchang Zheng
Journal:  Biomed Res Int       Date:  2017-07-12       Impact factor: 3.411

2.  Validation of inflationary noninvasive blood pressure monitoring in the emergency room.

Authors:  Junichi Sasaki; Yoshiharu Kikuchi; Takashi Usuda; Shingo Hori
Journal:  Blood Press Monit       Date:  2015-12       Impact factor: 1.444

3.  An Ambulatory Blood Pressure Monitor Mobile Health System for Early Warning for Stroke Risk: Longitudinal Observational Study.

Authors:  Guangyu Wang; Silu Zhou; Shahbaz Rezaei; Xin Liu; Anpeng Huang
Journal:  JMIR Mhealth Uhealth       Date:  2019-10-30       Impact factor: 4.773

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.