Noninvasive monitoring of stroke volume with resynchronization devices in patients with ischemic cardiomyopathy.

Michael Kühne1, Mario Bocchiardo, Herbert Nägele, Beat Schaer, Michael Lippert, Christian Sticherling, Stefan Osswald.   

Abstract

BACKGROUND: A novel method to estimate cardiac volumes based on impedance measurements using the leads of a resynchronization device has been developed. This study investigated the method in patients with ischemic cardiomyopathy and documented wall motion abnormalities. METHOD AND
RESULTS: Fifteen postinfarction patients (age 68 ± 8 years, ejection fraction 27 ± 5%) with symptomatic heart failure and ≥ 1 akinetic or dyskinetic segment were included. During the implantation of a cardiac resynchronization therapy (CRT) device, acute impedance curves were recorded along with stroke volume determined by the arterial pulse contour method. In an overdrive protocol, the impedance parameter "stroke impedance" decreased in significant correlation with stroke volume in all patients. The median correlation coefficient between stroke volume and stroke impedance was 0.83 (interquartile range 0.70-0.89). Furthermore, the atrioventricular delay was optimized based on impedance and reference stroke volume. After optimization by the impedance method, it differed by 18 ± 15 ms from the figure after optimization by the invasive reference. Compared with a standard atrioventricular delay of 120 ms, stroke volume was improved by 8.6 ± 9.8% with the use of invasive optimization and by 6.4 ± 10.8% with the use of impedance-based optimization.
CONCLUSIONS: In CRT patients with chronic infarction and wall motion abnormalities, impedance is a valid parameter to estimate stroke volume and to guide optimization of CRT timing.
Copyright © 2013 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Heart failure; bioelectrical impedance; cardiac output; cardiac resynchronization therapy; implantable defibrillators; patient monitoring

Mesh:

Year:  2013        PMID: 23910588     DOI: 10.1016/j.cardfail.2013.06.005

Source DB:  PubMed          Journal:  J Card Fail        ISSN: 1071-9164            Impact factor:   5.712


  2 in total

Review 1.  Recent advances in the optimization of cardiac resynchronization therapy.

Authors:  Satish Chandraprakasam; Gina G Mentzer
Journal:  Curr Heart Fail Rep       Date:  2015-02

2.  Hemodynamic monitoring by intracardiac impedance measured by cardiac resynchronization defibrillators: Evaluation in a controlled clinical setting (BIO.Detect HF II study).

Authors:  Peter-Paul Henri Marie Delnoy; Klaus-Jürgen Gutleben; Niels Eske Bruun; Sebastian K G Maier; Hanno Oswald; Christoph Stellbrink; Jens Brock Johansen; Stefan Paule; Peter Søgaard
Journal:  Indian Pacing Electrophysiol J       Date:  2021-04-15
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.