Experimental and Monte Carlo evaluation of Eclipse treatment planning system for effects on dose distribution of the hip prostheses.

Serap Catlı1, Güneş Tanır.   

Abstract

The present study aimed to investigate the effects of titanium, titanium alloy, and stainless steel hip prostheses on dose distribution based on the Monte Carlo simulation method, as well as the accuracy of the Eclipse treatment planning system (TPS) at 6 and 18MV photon energies. In the present study the pencil beam convolution (PBC) method implemented in the Eclipse TPS was compared to the Monte Carlo method and ionization chamber measurements. The present findings show that if high-Z material is used in prosthesis, large dose changes can occur due to scattering. The variance in dose observed in the present study was dependent on material type, density, and atomic number, as well as photon energy; as photon energy increased back scattering decreased. The dose perturbation effect of hip prostheses was significant and could not be predicted accurately by the PBC method for hip prostheses. The findings show that for accurate dose calculation the Monte Carlo-based TPS should be used in patients with hip prostheses.
Copyright © 2013 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Hip prostheses; Monte Carlo method; Radiotherapy; Treatment planning system

Mesh:

Year:  2013        PMID: 23910447     DOI: 10.1016/j.meddos.2013.03.005

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482


  5 in total

Review 1.  High-density dental implants and radiotherapy planning: evaluation of effects on dose distribution using pencil beam convolution algorithm and Monte Carlo method.

Authors:  Serap Çatli
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

2.  Dosimetric Evaluation of the QFix kVueTM Calypso Couch Top.

Authors:  Lingtong Hou; Huiqin Zhang; Xiaomei Sun; Qianqian Liu; Tingfeng Chen; Yong Liu; Xiaodong Jiang; Shengyu Yao
Journal:  Technol Cancer Res Treat       Date:  2021 Jan-Dec

3.  Radiotherapy-induced toxicity in prostate cancer patients with hip prostheses.

Authors:  Andrea M Fischer; Peter J Hoskin
Journal:  Radiat Oncol       Date:  2022-01-17       Impact factor: 3.481

4.  Monte Carlo modeling of the Elekta Versa HD and patient dose calculation with EGSnrc/BEAMnrc.

Authors:  Holly M Parenica Paschal; Christopher N Kabat; Pavlos Papaconstadopoulos; Neil A Kirby; Pamela A Myers; Timothy D Wagner; Sotirios Stathakis
Journal:  J Appl Clin Med Phys       Date:  2022-08-19       Impact factor: 2.243

5.  VMAT Optimization and Dose Calculation in the Presence of Metallic Hip Prostheses.

Authors:  Holly M Parenica; Panayiotis Mavroidis; William Jones; Gregory Swanson; Niko Papanikolaou; Sotirios Stathakis
Journal:  Technol Cancer Res Treat       Date:  2019 Jan-Dec
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.