High frequency alternating current chip nano calorimeter with laser heating.

E Shoifet1, Y Z Chua, H Huth, C Schick.   

Abstract

Heat capacity spectroscopy at frequencies up to 100 kHz is commonly performed by thermal effusivity measurements applying the 3ω-technique. Here we show that AC-calorimetry using a thin film chip sensor allows for the measurement of frequency dependent heat capacity in the thin film limit up to about 1 MHz. Using films thinner than the thermal length of the thermal wave (~1 μm) at such frequencies is advantageous because it provides heat capacity alone and not in combination with other quantities like thermal conductivity, at least on a qualitative basis. The used calorimetric sensor and the sample are each less than 1 μm thick. For high frequency AC-calorimetry, high cooling rates at very small temperature differences are required. This is realized by minimizing the heated spot to the size of the on chip thermocouple (3 × 6 μm(2)). A modulated laser beam shaped and positioned by a glass fiber is used as the heat source. The device was used to measure the complex heat capacity in the vicinity of the dynamic glass transition (structural relaxation) of poly(methyl methacrylate). Combining different calorimeters finally provides data between 10(-3) Hz and 10(6) Hz. In this frequency range the dynamic glass transition shifts about 120 K.

Entities:  

Year:  2013        PMID: 23902080     DOI: 10.1063/1.4812349

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523


  1 in total

1.  Adsorption Kinetics of Polystyrene and Poly(9-anthracenyl methyl methacrylate) onto SiO2 Surface Measured by Chip Nano-Calorimetry.

Authors:  Mina Ishihara; Tomoya Watanabe; Takashi Sasaki
Journal:  Polymers (Basel)       Date:  2022-02-03       Impact factor: 4.329

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.