In the proposed procedure for the determination of selenium, 0.5-1 g of sample is wet ashed with concentrated nitric acid. After adding 1,2-diamino-4,5-dichlorobenzene to the digest at pH 1, the resulting dichloropiazselenol derivatives is extracted with toluene. The extract is purified by column chromatography over Florisil and analyzed by gas-liquid chromatography with electron capture detection. Recoveries of selenium added to various substrates ranged from 72 to 102%. The limit of detection is approximately 0.01 ppm, but smaller amounts can be determined by increasing the sample size or by concentration of the final extract.

Entities:  

Mesh:

Substances:

Year:  1975        PMID: 239017     DOI: 10.1016/s0021-9673(01)91795-9

Source DB:  PubMed          Journal:  J Chromatogr


  1 in total

Review 1.  Determination of selenium in the environment and in biological material.

Authors:  E M Bem
Journal:  Environ Health Perspect       Date:  1981-01       Impact factor: 9.031