Literature DB >> 23900998

Collapse of the Advanta V12 Large Diameter covered stent following implantation for aortic coarctation.

Nicholas Hayes1, Tomaž Podnar, Shakeel Qureshi.   

Abstract

We report a case of collapse of the Advanta V12 Large Diameter (LD) covered stent following previous successful implantation for native aortic coarctation. This unreported complication was successfully managed with subsequent stent redilation and implantation of two covered Cheatham-Platinum stents within the collapsed Advanta V12 stent. This case highlights the importance of correct stent apposition to the aortic wall and also raises some concerns regarding the radial strength of the Advanta V12 stent.
Copyright © 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  cardiac catheterization; coarctation of aorta; stent

Mesh:

Year:  2013        PMID: 23900998     DOI: 10.1002/ccd.25139

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692