Literature DB >> 23897107

Studies on the use of BD Vacutainer® SST II™ and RST™ in general practice: investigation of artefactual hyperkalaemia.

Tine Huyghe1, Frank Buntinx, Rudi Bruyninckx, Veerle Besard, Jos Vunckx, Stephen Church, Karen Byron, Renee Rosa, Norbert Blanckaert.   

Abstract

BACKGROUND: Current sampling and transport conditions of samples in general practice can result in pseudohyperkalaemia. This study was undertaken to determine, in a general practice setting, whether there is any difference in haemolysis obtained when using BD Vacutainer® Rapid Serum Tubes (BD RST) compared with using BD Vacutainer® SST™ II Advance Blood Collection Tubes (BD SSTII).
METHODS: Blood was collected from 353 patients requiring blood sampling who were attending 31 general practitioner practices in Belgium. For each patient, two BD SSTII tubes and two BD RST tubes were drawn in a randomized order. One of each pair of tubes was inverted five times, the other was not. Serum potassium concentration, serum LDH activity and haemolysis index were measured in each sample.
RESULTS: There was no significant difference in measured potassium concentration according to tube type (P = 0.16). Measured LDH activities were 1.7% higher in serum collected into BD SSTII tubes compared to BD RST tubes (P = 0.02). When comparing serum from unmixed BD RST with BD SSTII tubes, there was a slight reduction in the haemolysis index but no significant difference in measured potassium concentration or LDH activity. Risk of hyperkalaemia was 4.8 times higher in serum from tubes that were incompletely filled compared to those that were filled with the correct amount of blood.
CONCLUSION: Both types of blood tubes are suitable for the measurement of serum potassium and LDH in patients from general practice. Tube inversion does not improve the accuracy of either serum potassium or LDH measurement. Blood tubes should be filled to the level recommended by the manufacturer to avoid artefactual increases in measured serum potassium concentration and LDH activity.

Entities:  

Keywords:  Electrolytes; analytes; clinical studies; evaluation of new methods; laboratory methods

Mesh:

Substances:

Year:  2013        PMID: 23897107     DOI: 10.1177/0004563213488758

Source DB:  PubMed          Journal:  Ann Clin Biochem        ISSN: 0004-5632            Impact factor:   2.057


  1 in total

1.  Evaluation of the BD Vacutainer(®) RST blood collection tube for routine chemistry analytes: clinical significance of differences and stability study.

Authors:  Marija Kocijancic; Jelena Cargonja; Alma Delic-Knezevic
Journal:  Biochem Med (Zagreb)       Date:  2014-10-15       Impact factor: 2.313

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.