Aerodynamic dose emission characteristics of dry powder inhalers using an Andersen Cascade Impactor with a mixing inlet: the influence of flow and volume.

Sani Ibn Yakubu1, Khaled H Assi, Henry Chrystyn.   

Abstract

An interaction between device resistance and inhalation flow provides the 'energy' to de-aggregate the metered dose of dry powder inhalers (DPIs). Hence all dry powder inhalers demonstrate flow dependent dose emission but information on this at low flows is not available. We have adapted the compendial method for the Andersen Cascade Impactor (ACI) to include a mixing inlet to determine the aerodynamic dose emission characteristics of a salbutamol Diskus(®) [DSK], Easyhaler(®) [EASY] and Clickhaler(®) [CLICK] and the terbutaline Turbuhaler(®) [TBH] using flows of 10-60 L/min and inhalation volumes of 2 and 4 L. All DPIs demonstrated flow dependent dose emission (p<0.001) but there was no difference in the measurements between 2 and 4 L. The flow dependent dose emission properties of each DPI started to plateau when the pressure change inside each device, during an inhalation, was between 1 and 1.5 kPa. This corresponds to inhalation flows of 40.1-49.1, 25.4-28.9, 23.6-28.9 and 29.7-36.3 L/min through DSK, CLICK, EASY and TBH. The adapted methodology allows measurements at low flows. The results highlight that the compendial methodology to use an inhaled volume of 4 L with the ACI could be replaced by 2 L and that the recommendation to make measurements using a pressure drop of 4kPa should be revised.
Copyright © 2013 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Different volumes; Dose emission; Dry powder inhaler; Low flows

Mesh:

Substances:

Year:  2013        PMID: 23892154     DOI: 10.1016/j.ijpharm.2013.07.036

Source DB:  PubMed          Journal:  Int J Pharm        ISSN: 0378-5173            Impact factor:   5.875


  4 in total

1.  The effect of inspiratory parameters after two separate inhalations on the dose emission of theophylline from low and high resistance dry powder inhalers.

Authors:  Mohamad Abadelah; Jumannah Al-Assadi; James Rooney; Hassane Larhrib
Journal:  Saudi Pharm J       Date:  2019-11-13       Impact factor: 4.330

2.  The contribution of patients' lung function to the inspiratory airflow rate achievable through a DPIs' simulator reproducing different intrinsic resistance rates.

Authors:  Roberto W Dal Negro; Paola Turco; Massimiliano Povero
Journal:  Multidiscip Respir Med       Date:  2021-04-15

Review 3.  Guiding Inspiratory Flow: Development of the In-Check DIAL G16, a Tool for Improving Inhaler Technique.

Authors:  Mark Jeremy Sanders
Journal:  Pulm Med       Date:  2017-11-16

4.  Lung Function Can Predict the Expected Inspiratory Airflow Rate through Dry Powder Inhalers in Asthmatic Adolescents.

Authors:  Roberto Walter Dal Negro; Paola Turco; Massimiliano Povero
Journal:  Children (Basel)       Date:  2022-03-08
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.