Accuracy of a digital impression system based on active wavefront sampling technology for implants considering operator experience, implant angulation, and depth.

Beatriz Giménez1, Mutlu Özcan, Francisco Martínez-Rus, Guillermo Pradíes.   

Abstract

BACKGROUND: There is a scarce knowledge on the accuracy of intraoral digital impression systems for dental implants.
PURPOSE: The purpose of this study is to evaluate the accuracy of a digital impression system considering clinical parameters.
MATERIALS AND METHODS: A master model with six implants (27, 25, 22, 12, 15, 17) was fitted with polyether ether ketone scan bodies. Implant no. 25 was placed with 30° mesial angulation in relation to the vertical plane (y axis), and implant no. 15 was positioned with 30° distal angulation. Implant no. 22 was placed 2 mm and no. 12, 4 mm below the gingiva. Experienced (n = 2) and inexperienced operators (n = 2) performed scanning (Lava Chairside Oral Scanner; 3 M ESPE, St Paul, MN, USA) at standard and high accuracy mode. Measurements involved five distances (27-25, 27-22, 27-12, 27-15, 27-17). Measurements with high accuracy three-dimensional coordinated measuring machine (CMM) of the master model acted as the true values. The data obtained were subtracted from those of the CMM values.
RESULTS: Experience of the operator significantly influenced the results (p = .000). Angulation (p = .195) and depth of implant (p = .399) did not show significant deviation from the true values. The mean difference between standard and high accuracy mode was 90 μm.
CONCLUSIONS: With the active wavefront sampling, technology-based digital impression system training seems to be compulsory. Impressions of angulated implants may diminish the accuracy of the impression, yet the results were not significant.
© 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  accuracy; dental implant; digital impression; implant angulation; implant depth; intraoral scanner

Mesh:

Substances:

Year:  2013        PMID: 23879869     DOI: 10.1111/cid.12124

Source DB:  PubMed          Journal:  Clin Implant Dent Relat Res        ISSN: 1523-0899            Impact factor:   3.932


  28 in total

1.  A new method for the evaluation of the accuracy of full-arch digital impressions in vitro.

Authors:  Jan-Frederik Güth; Daniel Edelhoff; Josef Schweiger; Christine Keul
Journal:  Clin Oral Investig       Date:  2015-10-10       Impact factor: 3.573

2.  Accuracy of full-arch digital impressions: an in vitro and in vivo comparison.

Authors:  Christine Keul; Jan-Frederik Güth
Journal:  Clin Oral Investig       Date:  2019-05-27       Impact factor: 3.573

3.  Accuracy of multi-unit implant impression: traditional techniques versus a digital procedure.

Authors:  Maria Menini; Paolo Setti; Francesco Pera; Paolo Pera; Paolo Pesce
Journal:  Clin Oral Investig       Date:  2017-09-30       Impact factor: 3.573

4.  In vitro assessment of the accuracy of digital impressions prepared using a single system for full-arch restorations on implants.

Authors:  Leonardo Ciocca; Roberto Meneghello; Carlo Monaco; Gianpaolo Savio; Lorenzo Scheda; Maria Rosaria Gatto; Paolo Baldissara
Journal:  Int J Comput Assist Radiol Surg       Date:  2018-03-02       Impact factor: 2.924

Review 5.  Precision and practical usefulness of intraoral scanners in implant dentistry: A systematic literature review.

Authors:  Ignacio García-Gil; Jorge Cortés-Bretón-Brinkmann; Jaime Jiménez-García; Jesus Peláez-Rico; María-Jesús Suárez-García
Journal:  J Clin Exp Dent       Date:  2020-08-01

6.  In Vitro Comparison of Three Intraoral Scanners for Implant-Supported Dental Prostheses.

Authors:  Vitória Costa; António Sérgio Silva; Rosana Costa; Pedro Barreiros; Joana Mendes; José Manuel Mendes
Journal:  Dent J (Basel)       Date:  2022-06-15

7.  Validity of Intraoral Scans Compared with Plaster Models: An In-Vivo Comparison of Dental Measurements and 3D Surface Analysis.

Authors:  Fan Zhang; Kyung-Jin Suh; Kyung-Min Lee
Journal:  PLoS One       Date:  2016-06-15       Impact factor: 3.240

Review 8.  Intraoral Scanner Technologies: A Review to Make a Successful Impression.

Authors:  Raphaël Richert; Alexis Goujat; Laurent Venet; Gilbert Viguie; Stéphane Viennot; Philip Robinson; Jean-Christophe Farges; Michel Fages; Maxime Ducret
Journal:  J Healthc Eng       Date:  2017-09-05       Impact factor: 2.682

Review 9.  Trueness and precision of digital implant impressions by intraoral scanners: a literature review.

Authors:  Minoru Sanda; Keita Miyoshi; Kazuyoshi Baba
Journal:  Int J Implant Dent       Date:  2021-07-27

10.  Bias Evaluation of the Accuracy of Two Extraoral Scanners and an Intraoral Scanner Based on ADA Standards.

Authors:  Naiyu Cui; Jiayin Wang; Xingyu Hou; Shixun Sun; Qixuan Huang; Ho-Kyung Lim; HongXin Cai; Qi Jia; Eui-Seok Lee; Heng Bo Jiang
Journal:  Scanning       Date:  2021-06-10       Impact factor: 1.932

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.