Clinical measurement of intravertebral pressure during vertebroplasty and kyphoplasty.

Yang Wang1, Fan Huang, Liang Chen, Zhen-Yong Ke, Zhong-Liang Deng.   

Abstract

BACKGROUND: Vertebroplasty (VP) and kyphoplasty (KP) are emerging procedures for almost immediate pain relief when treating osteoporotic or osteolytic fractures. The main reported complication is polymethylmethacrylate (PMMA) leakage, which may lead to compression of neural structures or embolism. Different authors have proposed that intravertebral pressure (IP) is an important factor determining the risk for leakage, although so far only limited information has been gathered from clinical and experimental studies. There is also a lack of understanding of the IP during conventional interventions in VP and KP in the clinic.
OBJECTIVE: (1) To compare the intravertebral pressures of compressed vertebrae and adjacent normal vertebrae. (2) To measure the IP of compressed vertebrae during VP and KP.
SETTING: An interventional pain management practice, a medical center, major metropolitan city, in the People's Republic of China.
METHODS: Thirty-five patients (with 40 compressed vertebrae and 35 adjacent normal vertebrae) were randomly allocated for intravertebral pressure measurements. Cannulas were placed bipedicularly into the posterior third of each vertebral body. Either PMMA or a balloon was injected into the vertebral body through the right cannula. A manometer was connected to the cannula in the left pedicle, and heparin was injected to verify the pressure measurement system.
RESULTS: The range (minimum-maximum), average IP, and the standard deviation of the compressed vertebrae were 0-39 mm Hg and 24.5 ±11.3 mm Hg; and that of adjacent normal vertebrae were 3-16 mm Hg, 7.3 ± 4.2 mm Hg. Furthermore, the average IP for Phase 1 (before PMMA injection) for VP was 23 ±11.9 mm Hg; the maximum IP recorded during injection was 169 ± 46.8 mm Hg and the IP for 10 minutes after injection was 33 ±9.4 mm Hg. Meanwhile, the highest IP recorded for KP patients was 142 ±39.6 mm Hg. The average IP for Phase 1 (before balloon inflation) was 24 ±12.7 mmHg; Phase 2 (peak IP during the balloon inflation) was 63 ± 25.8 mm Hg; and Phase 3 (after balloon inflation/before PMMA injection) was , and 18 ± 10.8 mm Hg. The IP for 10 minutes after injection in KP patients was 36 ± 8.5 mm Hg. LIMITATIONS: The flow rate was manually controlled, which is in line with clinical routine, and was kept at approximately 0.1 mL/s. Because the speed of injection was controlled by hand, an exact injection rate could not be assured, leading to some inaccuracy when comparing the IP of VP and KP patients. Each patient was injected with a different PMMA volume. Because PMMA injection was performed to a satisfactory vertebral body filling and limited by any signs of extravasation, it was difficult to maintain a constant injection volume, unlike in vitro studies. Other factors such as the damage to the vertebral shell or the degree of osteoporosis might also have affected the intravertebral pressure.
CONCLUSION: This study showed that the IP of compressed vertebrae was significantly higher than that of adjacent normal vertebrae. There was a significant increase in IP during the PMMA filling in VP and KP; the IP of compressed vertebrae was not significantly reduced by the balloon inflation in KP, and no statistically significant differences in IP were found during all common stages of PMMA filling in VP and KP.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23877465

Source DB:  PubMed          Journal:  Pain Physician        ISSN: 1533-3159            Impact factor:   4.965


  3 in total

1.  Anaphylactic shock during cement implantation of a total hip arthroplasty in a patient with underlying mastocytosis: case report of a rare intraoperative complication.

Authors:  Anita Ten Hagen; Pieter Doldersum; Tom van Raaij
Journal:  Patient Saf Surg       Date:  2016-11-05

2.  Risk factors for cement leakage and nomogram for predicting the intradiscal cement leakage after the vertebra augmented surgery.

Authors:  Tian-Yu Zhang; Pei-Xun Zhang; Feng Xue; Dian-Ying Zhang; Bao-Guo Jiang
Journal:  BMC Musculoskelet Disord       Date:  2020-11-30       Impact factor: 2.362

3.  Safety and efficacy studies of kyphoplasty, mesh-container-plasty, and pedicle screw fixation plus vertebroplasty for thoracolumbar osteoporotic vertebral burst fractures.

Authors:  Yimin Li; Yunfan Qian; Guangjie Shen; Chengxuan Tang; Xiqiang Zhong; Shaoqi He
Journal:  J Orthop Surg Res       Date:  2021-07-06       Impact factor: 2.359

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.