Behrang H Hamadani1, John Roller, Brian Dougherty, Fiona Persaud, Howard W Yoon.
Abstract
An irradiance mode, absolute differential spectral response measurement system for solar cells is presented. The system is based on combining the monochromator-based approach of determining the power mode spectral responsivity of cells with an LED-based measurement to construct a curve representing the light-overfilled absolute spectral response of the entire cell. This curve can be used to predict the short-circuit current (I(sc)) of the cell under the AM 1.5 standard reference spectrum. The measurement system is SI-traceable via detectors with primary calibrations linked to the NIST absolute cryogenic radiometer. An uncertainty analysis of the methodology places the relative uncertainty of the calculated I(sc) at better than ±0.8%.Entities:
Year: 2013 PMID: 23872765 DOI: 10.1364/AO.52.005184