Detailed analysis of polymer response to delivery balloon expansion of drug-eluting stents versus bare metal stents.

Scott J Denardo1, Paul L Carpinone, David M Vock, James E Tcheng, Harry R Phillips, Bradley J Willenberg, Christopher D Batich, Carl J Pepine.   

Abstract

AIMS: We sought to describe the response of the polymer surface of drug-eluting stents (DES) to delivery balloon expansion, including quantitation of any resulting detached microparticles. METHODS AND
RESULTS: We expanded the US Food and Drug Administration (FDA)-approved first- and second-generation DES in a vacuum filtration system and used optical and scanning electron microscopy to image the polymer surface, filters and delivery balloons. DES were expanded under a range of conditions, from in vitro conditions used for FDA regulatory submissions to human in vivo conditions. Dispersive Raman spectroscopy was used for definitive identification of microparticles. All polymer surfaces were topographically disturbed over an average of 4.6%-100% of the surface area imaged. Disturbances ranged from deformation (including peeling) to complete delamination. The dimensions of detached microparticles were 2-350 μm. The extent and nature of surface disturbances and microparticles were primarily a function of polymer composition (p<0.001 for 8/10 disturbance types/locations) and were independent of expansion condition (p=0.100 to 0.989 for 9/10 disturbance types/locations).
CONCLUSIONS: Balloon expansion of first- and second-generation DES disturbs the polymer surface and can cause detachment of microparticles; each is functionally related to the specific polymer but not to expansion condition. Disturbance "roughness" and detached microparticles may contribute to DES limitations.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23872653     DOI: 10.4244/EIJV9I3A62

Source DB:  PubMed          Journal:  EuroIntervention        ISSN: 1774-024X            Impact factor:   6.534


  4 in total

1.  The Coronary Microcirculation in STEMI: The Next Frontier?

Authors:  R David Anderson; Carl J Pepine
Journal:  Eur Heart J       Date:  2015-09-25       Impact factor: 29.983

2.  Challenges in Patients with Diabetes: Improving Clinical Outcomes After Percutaneous Coronary Intervention Through EVOlving Stent Technology.

Authors:  Robert A Byrne; Shmuel Banai; Roisin Colleran; Antonio Colombo
Journal:  Interv Cardiol       Date:  2018-01

3.  Meeting the Unmet - The Cre8 Polymer-free Drug-eluting Stents Technology: Proceedings of a satellite symposium held at EuroPCR on May 20th - 23rd 2014 in Paris.

Authors:  Katrina Mountfort; Didier Carrié; Marco Valgimigli; Gennaro Sardella; Shmuel Banai; Rafael Romaguera; Pieter Stella
Journal:  Interv Cardiol       Date:  2014-08

Review 4.  Drug-coated balloon therapy in coronary and peripheral artery disease.

Authors:  Robert A Byrne; Michael Joner; Fernando Alfonso; Adnan Kastrati
Journal:  Nat Rev Cardiol       Date:  2013-11-05       Impact factor: 32.419

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.