A comparison of precision of fit, rotational freedom, and torque loss with copy-milled zirconia and prefabricated titanium abutments.

Marzieh Alikhasi1, Abbas Monzavi, Seyed Hossein Bassir, Roshanak Baghai Naini, Nasim Khosronedjad, Sareh Keshavarz.   

Abstract

PURPOSE: The present study sought to compare copy-milled zirconia abutments with standard prefabricated titanium abutments in terms of horizontal and vertical misfit, rotational freedom, and torque loss.
MATERIALS AND METHODS: Twenty regular-platform implants (Brånemark, Nobel Biocare) with a standard external-hexagon implant-abutment interface were divided into two groups. In one group, prefabricated titanium abutments (Easy Abutment, Nobel Biocare) were used; in the other group, custom-made zirconia abutments were fabricated using the Zirkonzahn milling system. Measurements of geometry, rotational freedom, and marginal discrepancy were performed using a video measuring machine. To measure torque loss, each abutment was torqued to 32 Ncm and then loosened with a digital torque wrench. The torque loss values were reported as a percentage of the applied torque.
RESULTS: Rotational freedom and horizontal misfit were significantly higher with the customized abutments. There were no statistically significant differences in vertical misfit or in mean percentage of torque loss between the two groups.
CONCLUSIONS: Within the limitations of this study, it could be concluded that copy-milled zirconia abutments do not fit as accurately as prefabricated titanium abutments. However, the less precise fit of customized zirconia abutments does not seem to affect their ability to maintain the applied torque.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23869357     DOI: 10.11607/jomi.2937

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  8 in total

1.  In vitro performance and fracture resistance of CAD/CAM-fabricated implant supported molar crowns.

Authors:  Martin Rosentritt; Sebastian Hahnel; Frank Engelhardt; Michael Behr; Verena Preis
Journal:  Clin Oral Investig       Date:  2016-07-01       Impact factor: 3.573

2.  Accuracy of 3D white light scanning of abutment teeth impressions: evaluation of trueness and precision.

Authors:  Jin-Hun Jeon; Hae-Young Kim; Ji-Hwan Kim; Woong-Chul Kim
Journal:  J Adv Prosthodont       Date:  2014-12-17       Impact factor: 1.904

3.  Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and external-hex implant connections.

Authors:  Hakimeh Siadat; Elaheh Beyabanaki; Niloufar Mousavi; Marzieh Alikhasi
Journal:  J Adv Prosthodont       Date:  2017-08-16       Impact factor: 1.904

4.  Influence of different restorative materials on the stress distribution in dental implants.

Authors:  Carlos-Eduardo Datte; João-Paulo-Mendes Tribst; Amanda-Maria-de Oliveira Dal Piva; Renato-Sussumu Nishioka; Marco-Antonio Bottino; Alexandre-Duarte M Evangelhista; Fabrício M de M Monteiro; Alexandre-Luiz-Souto Borges
Journal:  J Clin Exp Dent       Date:  2018-05-01

5.  Marginal Discrepancy of Single Implant-Supported Metal Copings Fabricated by Various CAD/CAM and Conventional Techniques Using Different Materials.

Authors:  Safoura Ghodsi; Marzieh Alikhasi; Nika Soltani
Journal:  Eur J Dent       Date:  2019-11-26

6.  Influence of Implant Connection, Abutment Design and Screw Insertion Torque on Implant-Abutment Misfit.

Authors:  Jorge Vélez; Jesús Peláez; Carlos López-Suárez; Rubén Agustín-Panadero; Celia Tobar; María J Suárez
Journal:  J Clin Med       Date:  2020-07-24       Impact factor: 4.241

7.  Passant Connection Screw of Dental Implants: An In Vitro SEM Preliminary Study.

Authors:  Gabriele Cervino; Antonino Germanà; Luca Fiorillo; Cesare D'Amico; Francesco Abbate; Marco Cicciù
Journal:  Biomed Res Int       Date:  2022-04-21       Impact factor: 3.246

8.  Accuracy of digital and conventional dental implant impressions for fixed partial dentures: A comparative clinical study.

Authors:  Agne Gedrimiene; Rimas Adaskevicius; Vygandas Rutkunas
Journal:  J Adv Prosthodont       Date:  2019-10-30       Impact factor: 1.904

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.