Literature DB >> 2385819

A prototype retractor system designed to minimize ischemic brain retractor injury: initial observations.

A J Waring1, C M Housworth, R M Voorhies, J R Douglas, C F Walker, S E Connolly.   

Abstract

The authors have developed and patented a neurosurgical retractor system incorporating an infrared emitter and detector that allows detection of cerebral pulsations. Gentle contact with the surface of cat brains shows cerebral pulsations that correlate with arterial pulse as well as mechanical ventilation. The amplitude of cerebral pulsations decreases with higher retraction pressure and disappears at approximately 20 mmHg. The pressure on the surface of the brain decreased 50% in 5 minutes even though the position of the retractor was maintained constant. The authors postulate that monitoring cerebral pulsation may prove useful in clinical neurosurgery with respect to avoiding excessive retraction, which causes brain damage.

Entities:  

Mesh:

Year:  1990        PMID: 2385819     DOI: 10.1016/0090-3019(90)90062-t

Source DB:  PubMed          Journal:  Surg Neurol        ISSN: 0090-3019


  2 in total

1.  Technical pitfalls in a porcine brain retraction model. The impact of brain spatula on the retracted brain tissue in a porcine model: a feasibility study and its technical pitfalls.

Authors:  R Thiex; F J Hans; T Krings; B Sellhaus; J M Gilsbach
Journal:  Neuroradiology       Date:  2005-09-01       Impact factor: 2.804

2.  Surgical management of skull base tumors.

Authors:  Leonardo Rangel-Castilla; Jonathan J Russin; Robert F Spetzler
Journal:  Rep Pract Oncol Radiother       Date:  2014-10-14
  2 in total