Yongjia Li1, Duan Zhao, Wouter A Serdijn.
Abstract
A continuous-time level-crossing analog-to-digital converter (LC-ADC) for biomedical applications is presented. When compared to uniform-sampling (US) ADCs LC-ADCs generate fewer samples for various sparse biomedical signals. Lower power consumption and reduced design complexity with respect to conventional LC-ADCs are achieved due to: 1) replacing the n-bit digital-to-analog converter (DAC) with a 1-bit DAC; 2) splitting the level-crossing detections; and 3) fixing the comparison window. Designed and implemented in 0.18 μm CMOS technology, the proposed ADC uses a chip area of 220 × 203 μm(2). Operating from a supply voltage of 0.8 V, the ADC consumes 313-582 nW from 5 Hz to 5 kHz and achieves an ENOB up to 7.9 bits.Entities:
Mesh:
Year: 2013 PMID: 23853297 DOI: 10.1109/TBCAS.2013.2254484