| Literature DB >> 23853296 |
Xin Yi1, Jun Jia, Simin Deng, Steve Guofang Shen, Qing Xie, Guoxing Wang.
Abstract
Patients suffering from facial paralysis are on the hazard of disfigurement and loss of vision due to loss of blink function. Functional-electrical stimulation (FES) is one possible way of restoring blink and other functions in these patients. A blink restoration system for uni-lateral facial paralyzed patients is described in this paper. The system achieves restoration of synchronized blink through processing the myoelectric signal of orbicularis oculi at the normal side in real-time as the trigger to stimulate the paralyzed eyelid. Design issues are discussed, including EMG processing, stimulating strategies and real-time artifact blanking. Two artifact removal approaches based on sample and hold and digital filtering technique are proposed and implemented. Finally, the whole system has been verified on rabbit models.Entities:
Mesh:
Year: 2013 PMID: 23853296 DOI: 10.1109/TBCAS.2013.2255051
Source DB: PubMed Journal: IEEE Trans Biomed Circuits Syst ISSN: 1932-4545 Impact factor: 3.833