This paper presents for the first time experimental results of a current-to-voltage converter that can be used for amperometric sensing of currents ranging from 1 pA to 1 muA. The design strategy is optimized to achieve low power levels and, hence, make the circuit suitable for use in a wearable or implantable sensor. The power reduction is mostly achieved by combining transistors operating in the weak inversion region with floating-gate metal-oxide semiconductor devices and three different gain settings. The power consumption under normal operation is 9.82 muW.Entities:
Year: 2009 PMID: 23853290 DOI: 10.1109/TBCAS.2009.2032095
Source DB: PubMed Journal: IEEE Trans Biomed Circuits Syst ISSN: 1932-4545 Impact factor: 3.833