Literature DB >> 23850735

Evaluation of a commercially available carbohydrate deficient transferrin kit to detect beta-2-transferrin in cerebrospinal fluid using capillary electrophoresis.

Karina Rodríguez-Capote1, Joceline Turner, Joseph Macri.   

Abstract

OBJECTIVE: The aim of this study was to evaluate the CEofix™ CDT Capillary Electrophoresis (CE) kit for the detection of β-2-tranferrin (β-2-Tf) in cerebrospinal fluid (CSF). DESIGN AND
METHOD: Evaluation was performed according to CLSI EP5-A and EP12-A guidelines.
RESULTS: The method resolved β-2-Tf from other Tf isoforms. The C50 for β-Tf was determined to be 0.46 mg/L. Neither hemoglobin nor bilirubin co-migrated with β-2-Tf.
CONCLUSIONS: The CDT kit can be used for detecting β-2-Tf in CSF by CE.
© 2013.

Entities:  

Keywords:  CSF rhinorrhea; Capillary electrophoresis; Cerebrospinal fluid leakage; Desialylated transferrin; Tau-protein; β-2-transferrin

Mesh:

Substances:

Year:  2013        PMID: 23850735     DOI: 10.1016/j.clinbiochem.2013.07.001

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  1 in total

1.  Tau protein as a possible marker of cerebrospinal fluid leakage in cerebrospinal fluid rhinorrhoea: A pilot study.

Authors:  Jean-Baptiste Oudart; Laure Zucchini; François-Xavier Maquart; Xavier Dubernard; Marc Labrousse; Géraldine Fiabane; Alexandra Quedreux; Fabien Litre; Laurent Ramont
Journal:  Biochem Med (Zagreb)       Date:  2017-08-28       Impact factor: 2.313

  1 in total