Literature DB >> 23830677

The effect of a coronoid prosthesis on restoring stability to the coronoid-deficient elbow: a biomechanical study.

Alia B Gray1, Bashar Alolabi, Louis M Ferreira, George S Athwal, Graham J W King, James A Johnson.   

Abstract

PURPOSE: The coronoid process has been recognized as a critical component in maintaining elbow stability. In the case of comminuted coronoid fractures, where repair is not possible or has failed, a prosthesis may be beneficial in restoring the osseous integrity of the elbow joint. The hypothesis of this in vitro biomechanical study was that a coronoid prosthesis would restore stability to the coronoid-deficient elbow.
METHODS: An anatomically shaped metallic coronoid prosthesis was designed and developed based on computed tomography-derived measurements and optimized to account for average cartilage thickness. Elbow kinematics and stability were determined for 8 cadaveric arms in active and passive elbow flexion in the varus, valgus, horizontal, and vertical positions using an elbow motion simulator. Varus-valgus angulation and internal-external rotation of the ulna relative to the humerus were quantified in the intact state, after collateral ligament sectioning and repair (control state), after a simulated 40% transverse coronoid fracture, and after implantation of the coronoid prosthesis.
RESULTS: Internal rotation of the ulna increased with a 40% coronoid fracture in the horizontal and varus positions. Increases in varus angulation after coronoid fracture were also observed in the horizontal and varus positions, during active and passive flexion, respectively. Following implantation of the coronoid prosthesis, elbow kinematics were restored similar to control levels in all elbow positions.
CONCLUSIONS: Our findings support our hypothesis that an anatomically shaped coronoid prosthesis would be effective in restoring stability to the coronoid-deficient elbow. CLINICAL RELEVANCE: This study provides evidence that the use of an anatomical implant restores stability to the coronoid-deficient elbow and rationale for further study and development of this method. For comminuted coronoid fractures, where repair is not possible or has failed, our research indicates that a prosthesis may be a feasible treatment option.
Copyright © 2013 American Society for Surgery of the Hand. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Coronoid; biomechanics; elbow; fracture; prosthesis

Mesh:

Year:  2013        PMID: 23830677     DOI: 10.1016/j.jhsa.2013.05.004

Source DB:  PubMed          Journal:  J Hand Surg Am        ISSN: 0363-5023            Impact factor:   2.230


  4 in total

1.  Olecranon tip osteoarticular autograft transfer for irreparable coronoid process fracture: a biomechanical study.

Authors:  Miguel A Ramirez; Jose M Ramirez; Brent G Parks; Michael A Tsai; Anand M Murthi
Journal:  Hand (N Y)       Date:  2015-06-03

Review 2.  [Elbow dislocation fractures].

Authors:  S Siebenlist; K F Braun
Journal:  Unfallchirurg       Date:  2017-07       Impact factor: 1.000

3.  Treatment of the coronoid process fractures with anteromedial approach: a case report.

Authors:  Hongwei Chen; Ziyang Wang; Jun Li; Biao Zhong; Xieyuan Jiang
Journal:  Int J Clin Exp Med       Date:  2015-10-15

4.  Reconstruction of coronoid process of the ulna: a literature review.

Authors:  Shanwen Zhao; Canjun Zeng; Song Yuan; Runguang Li
Journal:  J Int Med Res       Date:  2021-04       Impact factor: 1.671

  4 in total