| Literature DB >> 2382956 |
Abstract
An improved HPLC method for the measurement of serum creatinine is described. Separation is effected using a strong cation exchange column at pH 4.7. Analytical recovery, precision and linearity are satisfactory, and the method correlates well with a kinetic Jaffé procedure. The method provides a means of accurately measuring creatinine and may be of use in the investigation of interference in creatinine assays.Entities:
Mesh:
Substances:
Year: 1990 PMID: 2382956 DOI: 10.1177/000456329002700307
Source DB: PubMed Journal: Ann Clin Biochem ISSN: 0004-5632 Impact factor: 2.057