Literature DB >> 23822943

Intermittent pneumatic compression. A comparison of femoral vein velocity with five different devices.

D Warwick1, K Dewbury, A Forrester.   

Abstract

AIM: Different mechanical devices for thromboprophylaxis have different flow characteristics. A new device (Vadoplex) has been developed to provide a short impulse around the calf, a concept derived from the efficacy of foot impulse technology. New devices should be compared with existing devices to establish whether it has a comparable ability to augment venous velocity. Objectives of the study were to compare the venous velocity induced by the Vadoplex with established intermittent pneumatic calf and leg compressors (Covidien and Huntleigh).
METHODS: The venous velocity was established in ten healthy volunteers with standardised Duplex ultraonography of the common femoral vein. Measurements were taken at rest and on standing, with each device inactive and active.
RESULTS: The Vadoplex induced an increased femoral venous velocity at least as enhanced as established calf and full leg sleeves.
CONCLUSION: Calf impulse technology is an alternative to other systems in enhancing femoral vein blood flow, itself a surrogate for assumed thromboprophylactic effect.

Entities:  

Mesh:

Year:  2013        PMID: 23822943

Source DB:  PubMed          Journal:  Int Angiol        ISSN: 0392-9590            Impact factor:   2.789


  1 in total

1.  Neuromuscular electrostimulation viathe common peroneal nerve promotes lower limb blood flow in a below-kneecast: A potential for thromboprophylaxis.

Authors:  D J Warwick; A Shaikh; S Gadola; M Stokes; P Worsley; D Bain; A T Tucker; S D Gadola
Journal:  Bone Joint Res       Date:  2013-09-02       Impact factor: 5.853