Fundamental radiological and geometric performance of two types of proton beam modulated discrete scanning systems.

J B Farr1, F Dessy, O De Wilde, O Bietzer, D Schönenberg.   

Abstract

PURPOSE: The purpose of this investigation was to compare and contrast the measured fundamental properties of two new types of modulated proton scanning systems. This provides a basis for clinical expectations based on the scanned beam quality and a benchmark for computational models. Because the relatively small beam and fast scanning gave challenges to the characterization, a secondary purpose was to develop and apply new approaches where necessary to do so.
METHODS: The following performances of the proton scanning systems were investigated: beamlet alignment, static in-air beamlet size and shape, scanned in-air penumbra, scanned fluence map accuracy, geometric alignment of scanning system to isocenter, maximum field size, lateral and longitudinal field uniformity of a 1 l cubic uniform field, output stability over time, gantry angle invariance, monitoring system linearity, and reproducibility. A range of detectors was used: film, ionization chambers, lateral multielement and longitudinal multilayer ionization chambers, and a scintillation screen combined with a digital video camera. Characterization of the scanned fluence maps was performed with a software analysis tool.
RESULTS: The resulting measurements and analysis indicated that the two types of delivery systems performed within specification for those aspects investigated. The significant differences were observed between the two types of scanning systems where one type exhibits a smaller spot size and associated penumbra than the other. The differential is minimum at maximum energy and increases inversely with decreasing energy. Additionally, the large spot system showed an increase in dose precision to a static target with layer rescanning whereas the small spot system did not.
CONCLUSIONS: The measured results from the two types of modulated scanning types of system were consistent with their designs under the conditions tested. The most significant difference between the types of system was their proton spot size and associated resolution, factors of magnetic optics, and vacuum length. The need and benefit of mutielement detectors and high-resolution sensors was also shown. The use of a fluence map analytical software tool was particularly effective in characterizing the dynamic proton energy-layer scanning.

Entities:  

Mesh:

Year:  2013        PMID: 23822445     DOI: 10.1118/1.4807643

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  11 in total

1.  Impact of range shifter material on proton pencil beam spot characteristics.

Authors:  Jiajian Shen; Wei Liu; Aman Anand; Joshua B Stoker; Xiaoning Ding; Mirek Fatyga; Michael G Herman; Martin Bues
Journal:  Med Phys       Date:  2015-03       Impact factor: 4.071

2.  Experimental and Monte Carlo characterization of a dynamic collimation system prototype for pencil beam scanning proton therapy.

Authors:  Blake R Smith; Mark Pankuch; Daniel E Hyer; Wesley S Culberson
Journal:  Med Phys       Date:  2020-09-09       Impact factor: 4.071

3.  Ultra-high dose rate radiation production and delivery systems intended for FLASH.

Authors:  Jonathan Farr; Veljko Grilj; Victor Malka; Srinivasan Sudharsan; Marco Schippers
Journal:  Med Phys       Date:  2022-05-05       Impact factor: 4.506

4.  An investigation into the robustness of dynamically collimated proton therapy treatments.

Authors:  Blake R Smith; Daniel E Hyer; Wesley S Culberson
Journal:  Med Phys       Date:  2020-05-16       Impact factor: 4.071

5.  Pencil beam scanning dosimetry for large animal irradiation.

Authors:  Liyong Lin; Timothy D Solberg; Alexandro Carabe; James E Mcdonough; Eric Diffenderfer; Jenine K Sanzari; Ann R Kennedy; Keith Cengel
Journal:  J Radiat Res       Date:  2014-05-22       Impact factor: 2.724

6.  Comparing proton treatment plans of pediatric brain tumors in two pencil beam scanning nozzles with different spot sizes.

Authors:  John C Kralik; Liwen Xi; Timothy D Solberg; Charles B Simone; Liyong Lin
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

7.  A benchmarking method to evaluate the accuracy of a commercial proton monte carlo pencil beam scanning treatment planning system.

Authors:  Liyong Lin; Sheng Huang; Minglei Kang; Petri Hiltunen; Reynald Vanderstraeten; Jari Lindberg; Sami Siljamaki; Todd Wareing; Ian Davis; Allen Barnett; John McGhee; Charles B Simone; Timothy D Solberg; James E McDonough; Christopher Ainsley
Journal:  J Appl Clin Med Phys       Date:  2017-02-02       Impact factor: 2.102

8.  Beam-specific planning target volumes incorporating 4D CT for pencil beam scanning proton therapy of thoracic tumors.

Authors:  Liyong Lin; Minglei Kang; Sheng Huang; Rulon Mayer; Andrew Thomas; Timothy D Solberg; James E McDonough; Charles B Simone
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

9.  Validation and clinical implementation of an accurate Monte Carlo code for pencil beam scanning proton therapy.

Authors:  Sheng Huang; Minglei Kang; Kevin Souris; Christopher Ainsley; Timothy D Solberg; James E McDonough; Charles B Simone; Liyong Lin
Journal:  J Appl Clin Med Phys       Date:  2018-07-30       Impact factor: 2.102

10.  Innovations and the Use of Collimators in the Delivery of Pencil Beam Scanning Proton Therapy.

Authors:  Daniel E Hyer; Laura C Bennett; Theodore J Geoghegan; Martin Bues; Blake R Smith
Journal:  Int J Part Ther       Date:  2021-06-25
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.