Literature DB >> 23811721

Failure analysis of an abutment fracture on single implant restoration.

Chiharu Imakita1, Makoto Shiota, Yoko Yamaguchi, Shohei Kasugai, Noriyuki Wakabayashi.   

Abstract

PURPOSE: This study evaluated the mechanism of an abutment fracture by means of a case-based finite element (FE) stress analysis.
MATERIALS AND METHODS: A patient presented with a complaint of fracture of an implant-retained single posterior restoration that had been in place for 6 years. The retrieved restoration was examined on the original working cast and subjected to microscopic observation for fracture surface analysis. A case-based FE analysis was conducted to detect the location and magnitude of stress concentration in the implant structures.
RESULTS: The fracture occurred at the concave neck of the abutment, with an evidence of fatigue fracture on the titanium surface. Ratchet marks were shown at the distal lingual edge, indicating that the crack proceeded from distolingual to mesial direction. The highest maximum tensile stress at the distolingual concave neck of the abutment was shown when the load was directed on the internal and external surfaces of the mesial buccal cusp.
CONCLUSIONS: It is suggested that an offset loading because of the cantilever structure caused a high tensile stress on the distolingual edge, resulting in the fatigue fracture of the abutment.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23811721     DOI: 10.1097/ID.0b013e31829a16fd

Source DB:  PubMed          Journal:  Implant Dent        ISSN: 1056-6163            Impact factor:   2.454


  1 in total

1.  Improvement in Fatigue Behavior of Dental Implant Fixtures by Changing Internal Connection Design: An In Vitro Pilot Study.

Authors:  Nak-Hyun Choi; Hyung-In Yoon; Tae-Hyung Kim; Eun-Jin Park
Journal:  Materials (Basel)       Date:  2019-10-07       Impact factor: 3.623

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.