G Gambarini1, M Carrara, C Tenconi, N Mantaut, M Borroni, D Cutajar, M Petasecca, I Fuduli, M Lerch, E Pignoli, A Rosenfeld.
Abstract
MOSkin detectors were studied to perform real-time in vivo dose measurements in high dose rate prostate brachytherapy. Measurements were performed inside an urethral catheter in a gel phantom simulating a real prostate implant. Measured and expected doses were compared and the discrepancy was found to be within 8.9% and 3.8% for single MOSkin and dual-MOSkin configurations, respectively. Results show that dual-MOSkin detectors can be profitably adopted in prostate brachytherapy treatments to perform real-time in vivo dosimetry inside the urethra.Keywords: High dose rate; In vivo dosimetry; MOSFET; Prostate brachytherapy
Mesh:
Year: 2013 PMID: 23810727 DOI: 10.1016/j.apradiso.2013.06.001