H Sudahar1, P G G Kurup, V Murali, P Mahadev, J Velmurugan.
Abstract
The present study is to analyze the CyberKnife hypofractionated dose distribution of localized prostate cancer in terms of high-dose rate (HDR) brachytherapy equivalent doses to assess the degree of HDR brachytherapy resemblance of CyberKnife dose distribution. Thirteen randomly selected localized prostate cancer cases treated using CyberKnife with a dose regimen of 36.25Gy in 5 fractions were considered. HDR equivalent doses were calculated for 30Gy in 3 fractions of HDR brachytherapy regimen. The D5% of the target in the CyberKnife hypofractionation was 41.57 ± 2.41Gy. The corresponding HDR fractionation (3 fractions) equivalent dose was 32.81 ± 1.86Gy. The mean HDR fractionation equivalent dose, D98%, was 27.93 ± 0.84Gy. The V100% of the prostate target was 95.57% ± 3.47%. The V100% of the bladder and the rectum were 717.16 and 79.6mm(3), respectively. Analysis of the HDR equivalent dose of CyberKnife dose distribution indicates a comparable resemblance to HDR dose distribution in the peripheral target doses (D98% to D80%) reported in the literature. However, there is a substantial difference observed in the core high-dose regions especially in D10% and D5%. The dose fall-off within the OAR is also superior in reported HDR dose distribution than the HDR equivalent doses of CyberKnife. CrownEntities:
Keywords: CyberKnife; HDR brachytherapy; Hypofractionation; Localized prostate
Mesh:
Year: 2013 PMID: 23809693 DOI: 10.1016/j.meddos.2013.04.006
Source DB: PubMed Journal: Med Dosim ISSN: 1873-4022 Impact factor: 1.482