Laboratory performance in albumin and total protein measurement using a commutable specimen: results of a College of American Pathologists study.

Stanley F Lo1, W Greg Miller, Basil T Doumas.   

Abstract

CONTEXT: Discrepant results for serum constituents were observed among peer groups in the College of American Pathologists Comprehensive Chemistry Survey.
OBJECTIVES: To assess the performance of serum albumin and total protein measurement procedures and to evaluate the commutability of the conventional survey specimens.
DESIGN: A fresh frozen, off-the-clot serum sample was included along with 4 conventional survey specimens. The fresh frozen, off-the-clot serum sample was prepared in a manner expected to confer commutability with native clinical samples.
RESULTS: For the fresh frozen, off-the-clot serum sample, the mean values for 17 peer-groups were -0.07 to 0.32 g/dL from the bromocresol green albumin designated comparison method, whereas 4 VITROS (Ortho Clinical Diagnostics, Rochester, New York) peer groups differed by -0.29 to -0.37 g/dL (15 of 21 differences [71%] had P < .001). For bromocresol purple albumin methods, the mean differences from the designated comparison method from 8 peer groups were 0.25 to 0.47 g/dL (all had P < .001). For total protein methods, 23 peer group mean values were -0.07 to 0.15 g/dL from the reference measurement procedure (12 of 24 [50%] had P < .001). The Beckman (Fullerton, California) Synchron LX20 had a bias of -0.30 g/dL (P <.001). The commutability of the conventional specimens was acceptable for 23 of 24 bromocresol green method-material combinations (96%) and 13 of 16 bromocresol purple albumin method-material combinations (81%). All (100%) of the 36 method-material combinations had acceptable commutability for total protein.
CONCLUSIONS: One (2.2%) of the instrument systems (Synchron) using bromocresol green and none (0%) of the instrument systems using bromocresol purple had satisfactory total-error performance for albumin measurement. Differences in results between bromocresol green and bromocresol purple methods precluded using common reference intervals for interpreting results for serum albumin. Eight of 9 instrument systems (86.5%) had satisfactory total-error performance for total protein measurement.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23808463     DOI: 10.5858/arpa.2012-0152-CP

Source DB:  PubMed          Journal:  Arch Pathol Lab Med        ISSN: 0003-9985            Impact factor:   5.534


  2 in total

1.  Role of Janus kinase 3 in mucosal differentiation and predisposition to colitis.

Authors:  Jayshree Mishra; Raj K Verma; Gianfranco Alpini; Fanyin Meng; Narendra Kumar
Journal:  J Biol Chem       Date:  2013-09-17       Impact factor: 5.157

2.  Generating method-specific Reference Ranges - A harmonious outcome?

Authors:  Graham R Lee; Alison Griffin; Kieran Halton; Maria C Fitzgibbon
Journal:  Pract Lab Med       Date:  2017-07-05
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.