| Literature DB >> 2380566 |
M D Kramer1, M Müller-Bardorff, M M Simon, W Tilgen, E Schickel, D Petzoldt.
Abstract
We report on an ELISA procedure for the quantitative analysis of total human leukocyte (PMN)-elastase, i.e., the simultaneous determination of (i) free (non-complexed) PMN-elastase and of (ii) PMN-elastase that is complexed to alpha 1 proteinase inhibitor. Simultaneous detection of both forms of PMN-elastase was achieved using a monoclonal antibody that recognizes a PMN-elastase-specific epitope present on both the uncomplexed and the complexed form of the enzyme. The test system described is reliable, easy to perform and permits the determination of total PMN-elastase in complex biological fluids such as plasma or seminal fluid. It is to be expected that this test system will be useful for investigations of human PMN-elastase in biological specimens obtained from both normal and pathological conditions.Entities:
Mesh:
Substances:
Year: 1990 PMID: 2380566 DOI: 10.1016/0022-1759(90)90230-s
Source DB: PubMed Journal: J Immunol Methods ISSN: 0022-1759 Impact factor: 2.303