Literature DB >> 23804719

Polytetrafluoroethylene bicuspid pulmonary valve implantation: experience with 126 patients.

James Anthony Quintessenza1, Jeffrey Phillip Jacobs, Paul Jubeong Chai, Victor O Morell, Harald Lindberg.   

Abstract

This article reports our initial experience in 126 consecutive patients treated with placement of a surgically created polytetrafluoroethylene (PTFE) bicuspid pulmonary valve at The Congenital Heart Institute of Florida (CHIF). A bicuspid pulmonary valve is created with PTFE and sutured into the right ventricular outflow tract. PTFE bicuspid pulmonary valves were placed in 126 patients (age: range, 3.1-64.7 years, mean, 17.9 years; weight: range, 14.2-113.6 kg, mean, 55.4 kg). All patients had pulmonary insufficiency, pulmonary stenosis, or both, most commonly after previous repair of tetralogy of Fallot (71 patients). Follow-up was up to 8.3 years (range, 0-8.3 years, mean, 3.34 years). Operative mortality was 1 patient (0.8%). Late mortality was non-valve-related in 3 patients (2.4%). The initial 84 patents in this series received valves constructed from PTFE with 0.6-mm thickness. The next 42 patients received valves constructed from PTFE with 0.1-mm thickness. Six patients of 126 (4.8%) required replacement of the PTFE bicuspid pulmonary valve because of immobile and calcified leaflets. All 6 who required replacement of the PTFE bicuspid pulmonary valve initially received a valve constructed from porous 0.6-mm PTFE material. We currently use nonporous 0.1-mm PTFE, which does not allow cellular in-growth and thickening. Early echocardiographic follow-up of these valve leaflets made with 0.1-mm PTFE has demonstrated improved leaflet mobility and pliability and lower transvalvar gradients. PTFE bicuspid pulmonary valve implantation is safe and effective and demonstrates acceptable performance for the intermediate term. It is anticipated that using thinner 0.1-mm PTFE will improve valve function and durability. Long-term follow-up is necessary to determine the true value of this technique.

Entities:  

Keywords:  cardiac surgery; congenital heart disease; database; outcomes; pulmonary insufficiency; pulmonary stenosis; results of treatment; tetralogy of Fallot

Year:  2010        PMID: 23804719     DOI: 10.1177/2150135110361509

Source DB:  PubMed          Journal:  World J Pediatr Congenit Heart Surg        ISSN: 2150-1351


  8 in total

Review 1.  Natural Polymers in Heart Valve Tissue Engineering: Strategies, Advances and Challenges.

Authors:  Diana Elena Ciolacu; Raluca Nicu; Florin Ciolacu
Journal:  Biomedicines       Date:  2022-05-08

Review 2.  Evolution of pulmonary valve reconstruction with focused review of expanded polytetrafluoroethylene handmade valves.

Authors:  Te-I Chang; Kang-Hong Hsu; Shao-Jung Li; Min-Kai Chuang; Chi-Wen Luo; Yi-Jen Chen; Chung-I Chang
Journal:  Interact Cardiovasc Thorac Surg       Date:  2021-04-19

3.  Surgical management of chronic pulmonary regurgitation after relief of right ventricular outflow tract obstruction.

Authors:  Cheul Lee
Journal:  Korean Circ J       Date:  2012-01-31       Impact factor: 3.243

Review 4.  Engineering Efforts to Refine Compatibility and Duration of Aortic Valve Replacements: An Overview of Previous Expectations and New Promises.

Authors:  Stefano Rizzi; Sara Ragazzini; Maurizio Pesce
Journal:  Front Cardiovasc Med       Date:  2022-04-18

5.  Transcatheter pulmonary valve replacement by hybrid approach using a novel polymeric prosthetic heart valve: proof of concept in sheep.

Authors:  Ben Zhang; Xiang Chen; Tong-yi Xu; Zhi-gang Zhang; Xin Li; Lin Han; Zhi-yun Xu
Journal:  PLoS One       Date:  2014-06-13       Impact factor: 3.240

6.  Numerical and in-vitro experimental assessment of the performance of a novel designed expanded-polytetrafluoroethylene stentless bi-leaflet valve for aortic valve replacement.

Authors:  Guangyu Zhu; Munirah Binte Ismail; Masakazu Nakao; Qi Yuan; Joon Hock Yeo
Journal:  PLoS One       Date:  2019-01-30       Impact factor: 3.240

Review 7.  Materials and manufacturing perspectives in engineering heart valves: a review.

Authors:  F Oveissi; S Naficy; A Lee; D S Winlaw; F Dehghani
Journal:  Mater Today Bio       Date:  2019-12-05

8.  Short-term outcome of Polytetrafluoroethylene Membrane Valve versus Transannular Pericardial patch Reconstruction of Right Ventricular Outflow Tract in Tetralogy of Fallot: a Randomized Controlled Trial.

Authors:  Sanjib Rawat; Vivek Jaswal; Shyam Kumar Singh Thingnam; Harkant Singh; Sachin Mahajan; Reuben Lamiaki Kynta; Goverdhan Dutt Puri; Manoj Kumar Rohit
Journal:  Braz J Cardiovasc Surg       Date:  2021-02-01
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.