Abstract

AIM: To compare current practice of cuff size selection for noninvasive blood pressure measurement in a single-centre, tertiary-level neonatal intensive care unit (visual assessment of bladder width/limb length closest to 2/3) with common recommendations for appropriate cuff selection.
METHODS: Visual assessment of the appropriate cuff size ('2/3 rule') for upper arm, forearm and calf in 103 neonates (309 cuff selections) was compared with the following recommendations: (i) Method A - guidelines of the cuff manufacturer, (ii) Method B - cuff width/limb circumference ratio 0.44-0.60 and (iii) Method C - cuff width/limb length ratio closest to 0.66.
RESULTS: The upper arm cuff size was correctly chosen in 84% of cases (Method A), 43% (Method B) and 56% (Method C). The forearm cuff size was correctly chosen in 94% of cases (Method A), 68% (Method B) and 54% (Method C). The calf cuff size was correctly chosen in 96% of cases (Method A), 72% (Method B) and 63% (Method C).
CONCLUSION: The accuracy of selecting cuff size by visual assessment is low. Further research on accurate cuff selection for neonates, including at the forearm and calf, is warranted. ©2013 Foundation Acta Paediatrica. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  Blood pressure measurement; Newborn; Noninvasive; Oscillometric method

Mesh:

Year:  2013        PMID: 23799976     DOI: 10.1111/apa.12328

Source DB:  PubMed          Journal:  Acta Paediatr        ISSN: 0803-5253            Impact factor:   2.299


  3 in total

1.  Differences in Blood Pressure in Infants After General Anesthesia Compared to Awake Regional Anesthesia (GAS Study-A Prospective Randomized Trial).

Authors:  M E McCann; D E Withington; S J Arnup; A J Davidson; N Disma; G Frawley; N S Morton; G Bell; R W Hunt; D C Bellinger; D M Polaner; A Leo; A R Absalom; B S von Ungern-Sternberg; F Izzo; P Szmuk; V Young; S G Soriano; J C de Graaff
Journal:  Anesth Analg       Date:  2017-09       Impact factor: 5.108

Review 2.  Hemodynamic dysfunction in neonatal sepsis.

Authors:  Ashraf Kharrat; Amish Jain
Journal:  Pediatr Res       Date:  2021-11-24       Impact factor: 3.756

Review 3.  Novelty in hypertension in children and adolescents: focus on hypertension during the first year of life, use and interpretation of ambulatory blood pressure monitoring, role of physical activity in prevention and treatment, simple carbohydrates and uric acid as risk factors.

Authors:  Mirella Strambi; Marco Giussani; Maria Amalia Ambruzzi; Paolo Brambilla; Ciro Corrado; Ugo Giordano; Claudio Maffeis; Silvio Maringhin; Maria Chiara Matteucci; Ettore Menghetti; Patrizia Salice; Federico Schena; Pietro Strisciuglio; Giuliana Valerio; Francesca Viazzi; Raffaele Virdis; Simonetta Genovesi
Journal:  Ital J Pediatr       Date:  2016-07-16       Impact factor: 2.638

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.