Underdetection of ventricular fibrillation despite ICD testing and high sinus R wave.

Dirk Bastian1.   

Abstract

Apart from monitoring shock efficacy, proof of flawless detection of induced ventricular fibrillation (VF) is a decisive argument in favor of implantable cardioverter defibrillator (ICD) testing. On the other hand, it has been observed that undersensing of VF is extremely rare with good sensing of the intrinsic R wave of  ≥ 5-7 mV. The case presented here shows limitations in both argumentations: Neither optimal R wave sensing during sinus rhythm nor repeated ICD testing could rule out or predict multiple erroneous detections of clinical VF episodes. This must be taken into consideration in the current discussion on the necessity of defibrillation testing. Further optimization of sensing technology should be a focus in the development of modern ICD systems so as to improve the safety and efficacy of ICD therapy.

Entities:  

Mesh:

Year:  2013        PMID: 23784201     DOI: 10.1007/s00399-013-0272-8

Source DB:  PubMed          Journal:  Herzschrittmacherther Elektrophysiol        ISSN: 0938-7412


  4 in total

1.  Undersensing of VF in a patient with optimal R wave sensing during sinus rhythm.

Authors:  Lukas R C Dekker; Tim A M Schrama; Frans H L Steinmetz; Raymond Tukkie
Journal:  Pacing Clin Electrophysiol       Date:  2004-06       Impact factor: 1.976

2.  Differences in the characteristics of induced and spontaneous episodes of ventricular fibrillation.

Authors:  Nigel A Lever; Emma G Newall; Peter D Larsen
Journal:  Europace       Date:  2007-09-14       Impact factor: 5.214

3.  Defibrillation threshold testing: tradition or necessity?

Authors:  Christof Kolb; Stylianos Tzeis; Bernhard Zrenner
Journal:  Pacing Clin Electrophysiol       Date:  2009-05       Impact factor: 1.976

Review 4.  The dilemma of ICD implant testing.

Authors:  Charles D Swerdlow; Andrea M Russo; Paul J Degroot
Journal:  Pacing Clin Electrophysiol       Date:  2007-05       Impact factor: 1.976

  4 in total
  1 in total

1.  Undetected ventricular fibrillation in a single-chamber implantable cardioverter-defibrillator: When the far-field channel sees more than the intraventricular channel.

Authors:  Milena Kubickova; Pavel Cervinka; Andreas Kucher; Zdenek Nastoupil; Roland X Stroobandt
Journal:  HeartRhythm Case Rep       Date:  2016-04-19
  1 in total