Stability indicating reverse phase high performance liquid chromatography method for the estimation of capsaicin.

Alankar Shrivastava1, Prachi Saxena.   

Abstract

BACKGROUND: Stability Indicating Method (SIM) is a quantitative analytical procedure used to detect a decrease in the amount of the active pharmaceutical ingredient (API) present due to degradation. According to Food and Drug Administration (FDA) guidelines, an SIM is defined as a validated analytical procedure that accurately and precisely measures active ingredients (drug substance or drug product) free from potential interferences, such as degradation products, process impurities, excipients, or other potential impurities, and the FDA recommends that all assay procedures for stability studies be stability indicating. Here in this study we developed simple precise and accurate stability indicating reverse-phase high-performance liquid chromatographic (RP-HPLC) assay to analyze capsaicin (CAP) at concentrations from 70 to 130 μg/mL.
MATERIALS AND METHODS: HPLC equipped with Photo diode Array (PDA) detector, Pump model 600E of Waters and Empower software. Column used for the separation is ODS (250 × 4.6 mm) with particle size 5 μm and all the reagents and water were of HPLC grade.
RESULTS: The chromatographic separation was carried out using mobile phase Acetonitrile (ACN):water:buffer::75:10:15 with a flow rate of 1 mL/min on a C18 column. The concentration of the eluting compounds was monitored by a UV detector at 280 nm. No interferences were observed when stress conditions were applied and analyzed. Linearity was established using visual method, residuals plot, Dixon, and lack of fitness test. Limit of detection and limit of quantification was found to be 52.9 and 160 ng/mL, respectively. Recovery studies prove this method as useful in recovering the analytes. Relative Standard Deviation (RSD) of inter- and intraday precision within the acceptable limit of 2% proves that this method is precise. No degradation was found with alkaline conditions, thermal and photodegradation.
CONCLUSIONS: This study can be used for successful separation of CAP and its potential degradants in bulk and formulations. UV spectra of one of the degradants are also presented here and can be the basis to generate chemistry of potential degradants when CAP is kept under environmental conditions.

Entities:  

Keywords:  Capsaicin; method validation; stability indicating method; stress conditions

Year:  2011        PMID: 23781444      PMCID: PMC3658042          DOI: 10.4103/2229-4708.84451

Source DB:  PubMed          Journal:  Pharm Methods        ISSN: 2229-4708


INTRODUCTION

Capsaicin (CAP) is the main alkaloid responsible for pungency in chillies and is used as a counterirritant balm for external application and it is also used in creamy to provide external pain relief for arthritis patients.[1] To date, research has shown that capsaicinoids, and CAP in particular, have a wide variety of biological and physiological activities which provide them functions, such as antioxidants, anticarcinogenics, promotion of energy metabolism and suppression of fat accumulation, and inflammations. However, the potential applications of these molecules are limited by the irritation caused by their pungency.[2] Capsaicin (trans-8-methyl-N-vanillyl-6-nonenamide) is a crystalline, lipophilic, colorless, and odorless alkaloid with the molecular formula C18H27NO3 [Figure 1]. Its molecular weight is 305.40 g/mol, and it is fat-, alcohol- and oil-soluble. CAP displays cis/trans isomerism because the double bond prevents internal rotation. CAP is always found as the trans isomer because in the cis form, the –CH(CH3)2 and the longer chain on the other side of the double bond will be close together, causing them to repel each other slightly. This additional strain imposed causes the cis isomer to be a less stable arrangement than the trans isomer.[2]
Figure 1

Chemical structure of capsaicin

Chemical structure of capsaicin CAP is known to be effectively absorbed topically from the skin. In a study of 12 subjects topically administered, a 3% CAP solution in three different vehicles (70% isopropyl alcohol; mineral oil; and propylene glycol in 20% alcohol), CAP was shown to be rapidly absorbed and to quickly reach maximum concentration when CAP is applied topically. The half-life of CAP was approximately 24 h.[3]. In an extensive study of CAP distribution in tissues, elimination and other active principles in animals with oral administration, nearly 94% of orally administered CAP was absorbed and maximum concentration in the blood was reached 1 h after administration. In addition, a maximum distribution of 24.4% of administered CAP in blood, liver, kidney, and intestine was seen in 1 h and then diminished notably until being undetected after 4 days.[4] CAP metabolism is apparently similar in human, rat, and dog microsome. Three major metabolites have been identified for capsaicin: 16-hydroxycapsaicin, 17-hydroxycapsaicin, and 16,17-dihydrocapsaicin.[5] In vitro studies in human skin, CAP biotransformation was found to be slow and most CAP remained unchanged, whereas a small fraction was metabolized to vanillylamine and vanillyl acid. This suggests that cytochrome P450 enzymes participate minimally in CAP transformation in skin in comparison to their role in hepatic metabolism. CAP is mainly eliminated by the kidneys with a small untransformed proportion excreted in the feces and urine.[46] There are many methods for estimation of CAP, including estimation by colorimetry,[7] capillary gas chromatography,[8] HPLC–fluorescence detector in human serum,[9] Ultra Performance Liquid Chromatography (UPLC),[10] liquid chromatography with electrochemical detection,[11] HPLC–MS–TOF,[12] High Performance Thin Layer Chromatography (HPTLC) method coupled with Gas Chromatography and Mass Spectroscopy (GC–MS) detection.[13] CAP is now utilized in various formulations as drug because of its proven biological activity. Forced degradation studies to know about the interaction of this drug with different environmental conditions are proposed and can be utilized as one of the quality control methods for pharmaceutical industries. Hence the aim of the present work is to develop and validate a reverse-phase high-performance liquid chromatography (RP–HPLC) method for determination of CAP in bulk, including its forced degradation studies.

MATERIALS AND METHODS

Waters HPLC equipped with PDA detector, Pump model 600E and Empower software. Column used for the separation is ODS (250 × 4.6 mm) with particle size 5 μm. CAP pure was purchased from Ozone International, Mumbai, India. All solvents were of HPLC grade and procured from Merck Limited, Mumbai.

Preparation of 0.05 % v/v Orthophosphoric acid (buffer)

Ortho Phosphoric Acid (OPA) (0.5 mL) was transferred into a 10 mL volumetric flask, volume was made up with HPLC grade water to obtain a solution of 5% v/v and 1 ml of this was further diluted up to 100 mL volume with HPLC grade water to obtain a solution of 0.05 % v/v. The solution was filtered with 0.45 μm filter.

Preparation of standard stock solution

Pure CAP (50 mg) was weighed and dissolved in ACN up to 25 mL to get the stock solution of 2000 μg/mL. The solution was then filtered with 0. 45 μm syringe filter.

Preparation of 0.1 N HCl

HCl (0.85 mL) was diluted up to 100 mL to obtain a solution of 0.1 normality. The solution was filtered with 0.45 μm filter.

Preparation of 0.1 N NaOH

NaOH (400 mg) was weighed and dissolved in a 100 mL volumetric flask and diluted up to the mark to obtain the desired concentration. The solution was filtered with 0.4 μm filter.

Selection for wavelength of detection

CAP (50 mg) was weighed and dissolved in methanol in 25 mL volumetric flask volume was made up to the mark with methanol to get stock solution of concentration 2000 μg/mL. From stock 0.25 mL was taken into a 10 mL volumetric flask and diluted up to the mark to obtain a solution of strength 50 μg/mL. This dilution was scanned in the UV spectrophotometer and λmax of CAP 280 nm was selected as wavelength of detection. Figure 2 represents the graph of CAP using UV spectrophotometer.
Figure 2

UV spectra of capsaicin showing peak at 280 nm

UV spectra of capsaicin showing peak at 280 nm By analyzing UV spectra of CAP 280 nm (λmax > 254 nm of CAP) was selected as wavelength for detection.

Optimization of chromatographic conditions

Various combinations of methanol, ACN, and water were used to get symmetrical peak and system suitability parameters, such as capacity factor, tailing factor, and number of theoretical plates. Conditions were best optimized using mobile phase ACN:water:buffer::75:10:15 with a flow rate of 1 mL/min. Retention Time (RT) of peak was 6.14, plate count was 2180 (desirable is >2000), Tailing Factor (TF), was 1.25 (desirable is <1.5), and capacity factor was 5.15 (desirable is from 2 to 10). Figure 3 is the optimized chromatographic condition of CAP.
Figure 3

Optimized chromatographic condition by proposed method

Optimized chromatographic condition by proposed method

Preparation of calibration curve

Suitable aliquots of standard stock (2000 μg / mL) 0.35, 0.4, 0.45, 0.5, 0.55, 0.6, and 0.65 mL were transferred into 10 ml volumetric flask and finally diluted to the mark with ACN to obtain a series of solutions of 70, 80, 90, 100, 110, 120, and 130 μg/mL, respectively. The prepared solutions were filtered with 0.45 μm filter and 20 μL was injected for each dilution. Same procedure was repeated three times. The regression equation and regression coefficient from the mean of three determinations of the calibration curve was found to be y = 16442x – 91815 and 0.999, respectively. Overlay chromatogram and calibration curve are presented in Figures 4 and 5, respectively.
Figure 4

Overlay of calibration curve

Figure 5

Calibration curve of capsaicin by the proposed method

Overlay of calibration curve Calibration curve of capsaicin by the proposed method

Validation of the proposed method

Specificity

Specificity of the stability indicating method was established by separation of the principal peak with degradants during forced degradation. The stress conditions utilized were acid hydrolysis, alkaline hydrolysis, oxidation by peroxide, heat including both wet heat and dry heat, and photo degradation by UV light. Overall these studies were aimed to degrade 10%–30% of the drug. Overall summary of degradation studies are presented under Table 1
Table 1

Overall summary of degradation studies

Overall summary of degradation studies

Acidic hydrolysis

Stock solution (0.5 mL) of CAP was transferred to a 10 mL volumetric flask; 1 mL 0.5 N HCl was added and kept for 4 h in dark, then the volume was made up to 10 mL. The solution was then filtered and three replicates of 20 μL were injected in the HPLC system. Figures 6 and 7 represent chromatogram after acidic hydrolysis and UV spectra of degradant appearing in the chromatogram, respectively.
Figure 6

Chromatogram obtained by degradation of capsaicin with HCl

Figure 7

UV spectra of degradant obtained by acidic hydrolysis

Chromatogram obtained by degradation of capsaicin with HCl UV spectra of degradant obtained by acidic hydrolysis

RESULTS

The RT of main peak and degradant was 6.1 and 4.08, respectively. The resolution in the peak of CAP and degradant was 4.4 (desirable > 2). The plate count, TF, and capacity factor of both the peaks were also in limits. The percentage recovery of CAP was 75.07%.

Alkaline hydrolysis

Stock solution (0.5 mL of 2000 μg/mL) of CAP was transferred into a 10 mL volumetric flask, to this 1 mL 0.1 N NaOH was added and diluted to about 8 mL. The pH was adjusted to 6.8 with OPA and volume was made up to 10 mL with ACN. The solution was filtered and 20 μL was injected in three replicates. Results: No degradation was found with 0.1 N NaOH.

Oxidative degradation by peroxide

0.5 ml of standard stock solution was transferred in 10 mL volumetric flask, then 1 mL H2O2 (6%) was added and kept for 2 h in dark. Final volume was made up to the mark with ACN. The solution was filtered and 20 μL was injected in HPLC system. Result: The RT of CAP and degradant peak was at 6.14 and 3.49, respectively, and the resolution between the peaks is 5.65, plate count, TF, and capacity factor were also within desirable limits. The percentage recovery is 69.7%. The respective chromatogram is presented in Figure 8.
Figure 8

Chromatogram obtained by degradation of capsaicin with H2O2

Chromatogram obtained by degradation of capsaicin with H2O2

Degradation by UV light

Standard stock solution (0.5 mL) was transferred into a 10 mL volumetric flask, 4–5 mL of ACN was added and kept in UV chamber for 12 h. The final volume was made up to the mark, filtered and 20 μL was injected into the HPLC system. Result: No degradation was found by UV light.

Degradation by dry heat

CAP (10 mg) was weighed and transferred into a 10 mL volumetric flask and kept in hot air oven at 100°C for 2 h. Then, volume was made up to 10 mL with ACN and 1 mL of this solution was further diluted up to 10 mL to obtain 100 μg/mL solution, this was filtered and 20 μL was injected in HPLC system. Result: No degradation was found by dry heat.

Degradation by wet heat

Stock (0.5 mL) was transferred into a 10 mL volumetric flask, about 5 mL ACN was added and kept in water bath maintained at 100°C for 2 h. Then, final volume was made up to the mark, the solution was filtered and 20 μL was injected in HPLC system. Result: No degradation was found by wet heat. The degradation studies shows that the drug is degraded under stressed acidic and oxidative conditions. The RT of degradant formed and the resolution between main and degradant peak by acidic hydrolysis and oxidation was 4.08 and 3.49, and 4.4 and 5.65, respectively [Table 1]. Linearity is accessed by four methods:

Visualizing method

Out of seven, location of three points below, three above and one on the trend line in the calibration curve shows the linearity. Residual curve Result is mentioned in Figure 9.
Figure 9

Residual curve for capsaicin

Residual curve for capsaicin Dixon test for outliers: Data from Dixon test are presented in Tables 2 and 3.
Table 2

Ascending series of data of calibration curve

Table 3

Data from Dixon test for outliers

Ascending series of data of calibration curve Data from Dixon test for outliers Result: There are no outliers in the data of calibration curve according to Dixon test.

Lack-of-fitness test

Linear function analysis or lack-of-fitness test is applied by calculation of SSr, SSє, SSlof, and their respective variances. The applicability of the method was analyzed by comparing the tabulated and calculated F ratio. Data for lack-of-fitness test is presented in Table 4.
Table 4

Data for lack-of-fitness test for capsaicin

Data for lack-of-fitness test for capsaicin

Calculation of Error Sum Squares:

Residual error sum squares Lack-of-fit error sum squares Calculation of Degrees of freedom: DFr = (IJ – 2) = 19 DFє = (IJ – I) = 14 DFlof = (I – 2) = 5 Calculation of associated variance σr2 = SSr/DFr=1.47E+08 σє2 = SSr/DFє = 8.76E+06 σlof2 = SSlof/DFlof = 3.14E+0 Acceptability of linearity data F ratio = σlof2/σє2 = 3.59 Result: F tabulated at 95% confidence level is 4.56 and F calculated is 3.59, thus F (tabulated) > F(calculated), therefore the method is linear.

Range

70–130 μg/mL. 80–-120 μg/mL. –0.66–130 μg/mL. 100 μg/mL.

Precision

Repeatability: Repeatability was accessed by six replicates of test concentration, that is, 100 μg/mL; 20 μL was injected into the HPLC system. Intraday precision: 0.4, 0.5, and 0.6 mL were taken from the standard stock solution and diluted to 10 mL to obtain the dilution of 80, 100, and 120 μg/mL solutions, respectively. Three replicates were injected three times a day. Interday precision: Same procedure was followed to obtain 80%, 100%, and 120% of test concentration. Results: The RSD of repeatability was 0.298 (desirable < 1%) and RSD of intraday and interday precision was 0.517 and 0.810, respectively (desirable < 2%). Hence, the method is precise.

Accuracy

A sample of 70 μg/mL was prepared by diluting 0.35 mL standard stock to 10 mL with ACN. In three 10 mL volumetric flasks 0.35 mL of standard stock was transferred and to them 0.05, 0.15, and 0.25 mL standard stock was spiked, respectively. Each of them diluted up to the mark and filtered with 0.45 μm syringe filter. Three replicates of 20 μL of each sample were injected [Table 5].
Table 5

Recovery study of capsaicin

Recovery study of capsaicin Result: The method recovered 98.9%–100.9% (desirable 98%–102%) of the analyte. Hence, the method is accurate.

Limit of detection and limit of quantification

Limit of detection (LOD) and limit of quantification (LOQ) were calculated through standard deviation of the response of calibration curve and were found to be 52.9 and 160 ng/mL, respectively. LOD =3Sa/b LOQ = 10Sa/b where Sa is the standard deviation of the response and b is the slope of the calibration curve.[14]

System suitability testing

System suitability testing was performed by using six replicates of test concentrations. Variations in Tailing factor, asymmetry factor RT, and theoretical plates (N) were calculated as average of six replicates [Table 6].
Table 6

Summary of system suitability parameters

Summary of system suitability parameters Result: The RT, number of theoretical plates, capacity factor, tailing factor, and symmetry factor were found to be 6.14, 2179 (desirable > 2000), 5.26 (desirable 2–10), 1.25 (desirable < 1.5), and 1.25 (desirable < 2), respectively, from the mean of six determinations of test concentration.

CONCLUSIONS

In the present work, stability-indicating assay of CAP in bulk has been developed and validated. The method is capable of discriminating between the major active pharmaceutical ingredient from its degradation product formed by stressed conditions of acidic hydrolysis and oxidative degradation by peroxide. The developed method was validated in terms of specificity, linearity, precision, and accuracy. The method is specific as by forced degradation with acidic and oxidative conditions, the API peak was discriminating from the degradant peak with proper resolution. No degradation was found with alkaline conditions, thermal degradation, including both wet heat and dry heat and also with photostability. Linearity was proved by Dixon test of outliers and lack-of-fitness test. Limit of detection and limit of quantification was found to be 52.9 and 160 ng/mL, respectively; and recovery studies show that through this method, it is possible to recover the analyte. RSD of interday and intraday precision being within the acceptable limit of 2% proves that this method is precise. Results of degradation studies, summary of validation parameters, and optimized chromatographic condition; summary of validation parameters; and summary of system suitability parameters are presented in Tables 1, 7, and 8, respectively.
Table 7

The final optimized chromatographic conditions for the method

Table 8

Summary of validation parameters

The final optimized chromatographic conditions for the method Summary of validation parameters
  6 in total

1.  Determination of capsaicin and dihydrocapsaicin in capsicum fruits by liquid chromatography-electrospray/time-of-flight mass spectrometry.

Authors:  Ana Garcés-Claver; María S Arnedo-Andrés; Javier Abadía; Ramiro Gil-Ortega; Ana Alvarez-Fernandez
Journal:  J Agric Food Chem       Date:  2006-12-13       Impact factor: 5.279

2.  Determination of capsaicin and zucapsaicin in human serum by high-performance liquid chromatography with fluorescence detection.

Authors:  J Lu; M Cwik
Journal:  J Chromatogr B Biomed Sci Appl       Date:  1997-11-07

3.  Tissue distribution & elimination of capsaicin, piperine & curcumin following oral intake in rats.

Authors:  D Suresh; K Srinivasan
Journal:  Indian J Med Res       Date:  2010-05       Impact factor: 2.375

4.  Effects of vehicle on the uptake and elimination kinetics of capsaicinoids in human skin in vivo.

Authors:  Lynn K Pershing; Christopher A Reilly; Judy L Corlett; Dennis J Crouch
Journal:  Toxicol Appl Pharmacol       Date:  2004-10-01       Impact factor: 4.219

5.  In vitro hepatic and skin metabolism of capsaicin.

Authors:  Sanjay Chanda; Mohammad Bashir; Sunita Babbar; Aruna Koganti; Keith Bley
Journal:  Drug Metab Dispos       Date:  2008-01-07       Impact factor: 3.922

Review 6.  Chemical and pharmacological aspects of capsaicin.

Authors:  Maria de Lourdes Reyes-Escogido; Edith G Gonzalez-Mondragon; Erika Vazquez-Tzompantzi
Journal:  Molecules       Date:  2011-01-28       Impact factor: 4.411

  6 in total
  1 in total

1.  Curcumin and Capsaicin-Loaded Ag-Modified Mesoporous Silica Carriers: A New Alternative in Skin Treatment.

Authors:  Ivalina Trendafilova; Ralitsa Chimshirova; Denitsa Momekova; Hristo Petkov; Neli Koseva; Penka Petrova; Margarita Popova
Journal:  Nanomaterials (Basel)       Date:  2022-09-05       Impact factor: 5.719

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.