Literature DB >> 23773773

Incidence and the clinical impact of stent fractures after primary stenting for TASC C and D femoropopliteal lesions at 1 year.

J M Davaine1, J Quérat, B Guyomarch, M Á Brennan, A Costargent, P Chaillou, P Patra, Y Gouëffic.   

Abstract

BACKGROUND: The clinical impact of stent fractures is still controversial. This study analyzed the incidence and the clinical impact of stent fractures after stenting of long femoropopliteal lesions.
METHODS: From November 2008 to October 2009, 58 patients (62 limbs) were treated in a single center with a primary nitinol self-expanding stent for Trans-Atlantic Inter-Consensus (TASC) C and D de novo femoropopliteal lesions. Patients were prospectively followed by medical and duplex scan examinations. Stent fractures were assessed by biplane X-rays at 12 months. Logistic regression analysis was performed.
RESULTS: At 1 year a complete follow-up was obtained in 42 limbs/90 stents. The median length of the stented segment was 240 ± 180 cm with a mean of 2.1 (1-4) stents per patient. Sixteen stents (17.8%) were fractured: one type I (asymptomatic); seven type II (2 restenosis); five type III (asymptomatic), and three type IV stent fractures (1 restenosis). Stent diameter (p = .04) and stent implantation in the distal part of the superficial femoral artery (p = .05) were positively associated with stent fractures. Stent fracture had no influence on restenosis.
CONCLUSION: This study suggests that the high stent fracture rate associated with endovascular treatment of long femoropopliteal lesions should be balanced with its low clinical impact.
Copyright © 2013 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Clinical improvement; In-stent restenosis; Nitinol primary stenting; Patency; Stent fractures; Superficial femoral artery; TASC C and D lesions

Mesh:

Substances:

Year:  2013        PMID: 23773773     DOI: 10.1016/j.ejvs.2013.05.010

Source DB:  PubMed          Journal:  Eur J Vasc Endovasc Surg        ISSN: 1078-5884            Impact factor:   7.069


  6 in total

1.  Early results with LifeStent implantation in RESILIENT and non-RESILIENT inclusion criteria patients.

Authors:  Patrick A Stone; John E Campbell; Rashi Fischer; David Phang; Stephanie N Thompson; Neil Dippel; Albeir Mousa
Journal:  Vascular       Date:  2014-08-13       Impact factor: 1.285

Review 2.  Haemodynamics and Flow Modification Stents for Peripheral Arterial Disease: A Review.

Authors:  Efstratios Kokkalis; Nicolas Aristokleous; J Graeme Houston
Journal:  Ann Biomed Eng       Date:  2015-10-14       Impact factor: 3.934

3.  A case report of successful stent implantation through a fractured stent-strut in a superficial femoral artery based on bench testing simulation.

Authors:  Haruya Yamane; Shumpei Kosugi; Motoo Date; Yasunori Ueda
Journal:  Eur Heart J Case Rep       Date:  2021-07-01

4.  Prolonged high-pressure balloon angioplasty of femoropopliteal lesions: Impact on stent implantation rate and mid-term outcome.

Authors:  Gianluca Rigatelli; Mariano Palena; Paolo Cardaioli; Fabio dell'Avvocata; Massimo Giordan; Dobrin Vassilev; Marco Manzi
Journal:  J Geriatr Cardiol       Date:  2014-06       Impact factor: 3.327

5.  In vitro and in vivo Assessment of Keratose as a Novel Excipient of Paclitaxel Coated Balloons.

Authors:  Emily Turner; Megan Erwin; Marzieh Atigh; Uwe Christians; Justin M Saul; Saami K Yazdani
Journal:  Front Pharmacol       Date:  2018-07-30       Impact factor: 5.810

6.  Symptomatic superficial femoral artery pseudoaneurysm due to late stent fracture.

Authors:  Victor Bilman; Vincenzo Ardita; Alessandro Grandi; Roberto Chiesa; Luca Bertoglio
Journal:  J Vasc Surg Cases Innov Tech       Date:  2020-02-20
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.