Design and construction of a single tube, quantitative endpoint, LATE-PCR multiplex assay for ventilator-associated pneumonia.

L M Rice1, A H Reis, R Mistry, H Khan, P Khosla, S Bharya, L J Wangh.   

Abstract

AIMS: The goal of this study was to develop a molecular diagnostic multiplex assay for the quantitative detection of microbial pathogens commonly responsible for ventilator-associated pneumonia (VAP) and their antibiotic resistance using linear-after-the-exponential polymerase chain reaction (LATE-PCR). METHOD AND
RESULTS: This multiplex assay was designed for the quantitative detection and identification of pathogen genomic DNA of methicillin-resistant Staphylococcus aureus (MRSA), Acinetobacter baumannii, Pseudomonas aeruginosa, plus a control target from Lactococcus lactis. After amplification, the single-stranded amplicons were detected simultaneously in the same closed tube by hybridization to low-temperature molecular beacon probes labelled with four differently coloured fluorophores. The resulting hybrids were then analysed by determining the fluorescence intensity of each of the four fluorophores as a function of temperature.
CONCLUSIONS: This LATE-PCR single tube multiplex assay generated endpoint fluorescent contours that allowed identification of all microbial pathogens commonly responsible for VAP, including MRSA. The assay was quantitative, identifying the pathogens present in the sample, no matter whether there were as few as 10 or as many 100 000 target genomes. SIGNIFICANCE AND IMPACT OF THE STUDY: This assay is rapid, reliable and sensitive and is ready for preclinical testing using samples recovered from patients suffering from ventilator-associated pneumonia.
© 2013 The Society for Applied Microbiology.

Entities:  

Keywords:  LATE-PCR; VAP; endpoint; low-temperature molecular beacon probes; multiplex assay

Mesh:

Year:  2013        PMID: 23773171     DOI: 10.1111/jam.12281

Source DB:  PubMed          Journal:  J Appl Microbiol        ISSN: 1364-5072            Impact factor:   3.772


  2 in total

1.  A pilot study of the noninvasive assessment of the lung microbiota as a potential tool for the early diagnosis of ventilator-associated pneumonia.

Authors:  Addison K May; Jacob S Brady; Joann Romano-Keeler; Wonder P Drake; Patrick R Norris; Judith M Jenkins; Richard J Isaacs; Erik M Boczko
Journal:  Chest       Date:  2015-06       Impact factor: 9.410

2.  Development of two real-time multiplex PCR assays for the detection and quantification of eight key bacterial pathogens in lower respiratory tract infections.

Authors:  N J Gadsby; M P McHugh; C D Russell; H Mark; A Conway Morris; I F Laurenson; A T Hill; K E Templeton
Journal:  Clin Microbiol Infect       Date:  2015-05-14       Impact factor: 8.067