Comparison of two series of non-invasive instruments used for the skin physiological properties measurements: the 'Soft Plus' from Callegari S.p.A vs. the series of detectors from Courage & Khazaka.

W Hua1, H Xie, T Chen, L Li.   

Abstract

BACKGROUND: The invasive measuring techniques are of great help in the evaluation of skin physiological functions and topical formulation. The reproducibility and accuracy of the instrument are crucial. AIMS: The aim of this study is to reveal the differences and correlations in measuring skin hydration, pH, sebum, elasticity and melanin on the forearm or faces between two commercially available series of instruments. MATERIALS &
METHODS: 20 subjects were enrolled to be measured by the two series of instruments. The measurements by each series were performed on the left/right side of the body randomly. The pH, sebum, elasticity and melanin measurements were performed on the faces in different areas, while the hydration measurements were operated on the forearms before and after wet dressing, respectively.
RESULTS: Positive correlations were found in the values of skin SC hydration, pH, sebum and melanin detected by the two series. The values related to skin elasticity measured by the two instruments were statistically negative correlated. The Soft Plus series has higher variation in detecting skin hydration, pH, and melanin than the instruments from Courage & Khazaka. DISCUSSION: The differences of the two series must be instrument related. The difference in level of measurement can be due either to differences in the design of the probe, to differences in time for measurement, to differences in left/right part of the body or a combination of these three.
CONCLUSION: The two series have significant correlations. The instruments from CK are more reproducible.
© 2013 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  elasticity; melanin; non-invasive measurement; sebum; skin physiological properties; skin surface pH; stratum corneum hydration

Mesh:

Substances:

Year:  2013        PMID: 23772826     DOI: 10.1111/srt.12086

Source DB:  PubMed          Journal:  Skin Res Technol        ISSN: 0909-752X            Impact factor:   2.365


  2 in total

1.  Skin measurement devices to assess skin quality: A systematic review on reliability and validity.

Authors:  Mirte Langeveld; Lara S van de Lande; Eimear O' Sullivan; Berend van der Lei; Joris A van Dongen
Journal:  Skin Res Technol       Date:  2021-11-09       Impact factor: 2.240

2.  Objective and Patient-reported Assessments of Skin Grafts and Keystone Flaps-A Pilot Retrospective Cohort Study.

Authors:  Thomas D Dobbs; Thomas H Jovic; Zita M Jessop; Amanda Kyle; Hayley A Hutchings; Iain S Whitaker
Journal:  Plast Reconstr Surg Glob Open       Date:  2018-11-16
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.