PURPOSE: A new Monte Carlo (MC) source model (SM) has been developed for azimuthally symmetric photon beams.
METHODS: The MC simulation tallied phase space file (PSF) is divided into two categories depending on the relationship of the particle track line to the beam central axis: multiple point source (MPS) and spatial mesh based surface source (SMBSS). To validate this SM, MCNPX2.6 was used to generate two PSFs for a 6 MV photon beam from a Varian 2100C/D linear accelerator.
RESULTS: PDDs and profiles were calculated using the SM and original PSF for different field sizes from 5 × 5 to 40 × 40 cm2. Agreement was within 2% of the maximum dose at 100 cm SSD and 2.5% of the maximum dose at 200 cm SSD for beam profiles at depths of 3.5 cm and 15 cm with respect to the original PSF. Differences between the source model and the PSF in the out-of-field regions were less than 0.5% of the profile maximum value at 100 cm SSD. Differences between measured and calculated points were less than 2% of the maximum dose or 2 mm distance to agreement (DTA) at 100 cm SSD.
CONCLUSIONS: This SM is unique in that it accounts for a higher level of energy dependence on the particle's direction and it is independent from accelerator components, unlike other published SMs. The model can be applied to any arbitrary azimuthally symmetric beam and has source biasing capabilities that significantly increase the simulation speed up to 3300 for certain field sizes.
Copyright © 2013 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Monte Carlo; Phase space file; Radiotherapy; Source model

Mesh:

Year:  2013        PMID: 23768452     DOI: 10.1016/j.ejmp.2013.05.040

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  1 in total

1.  Spatial Mesh-Based Surface Source Model for the Electron Contamination of an 18 MV Photon Beams.

Authors:  Ahad Ollah Ezzati; Matthew T Studenski; Masuomeh Gohari
Journal:  J Med Phys       Date:  2021-02-02
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.