Literature DB >> 23761221

Benefits of smart pumps for automated changeovers of vasoactive drug infusion pumps: a quasi-experimental study.

M Cour1, R Hernu, T Bénet, J M Robert, D Regad, B Chabert, A Malatray, S Conrozier, P Serra, M Lassaigne, P Vanhems, L Argaud.   

Abstract

BACKGROUND: Manual changeover of vasoactive drug infusion pumps (CVIP) frequently lead to haemodynamic instability. Some of the newest smart pumps allow automated CVIP. The aim of this study was to compare automated CVIP with manual 'Quick Change' relays.
METHODS: We performed a prospective, quasi-experimental study, in a university-affiliated intensive care unit (ICU). All adult patients receiving continuous i.v. infusion of vasoactive drugs were included. CVIP were successively performed manually (Phase 1) and automatically (Phase 2) during two 6-month periods. The primary endpoint was the frequency of haemodynamic incidents related to the relays, which were defined as variations of mean arterial pressure >15 mm Hg or heart rate >15 bpm. The secondary endpoints were the nursing time dedicated to relays and the number of interruptions in care because of CVIP. A multivariate mixed effects logistic regression was fitted for analytic analysis.
RESULTS: We studied 1329 relays (Phase 1: 681, Phase 2: 648) from 133 patients (Phase 1: 63, Phase 2: 70). Incidents related to CVIP decreased from 137 (20%) in Phase 1 to 73 (11%) in Phase 2 (P<0.001). Automated relays were independently associated with a 49% risk reduction of CVIP-induced incidents (adjusted OR=0.51, 95% confidence interval 0.34-0.77, P=0.001). Time dedicated to the relays and the number of interruptions in care to manage CVIP were also significantly reduced with automated relays vs manual relays (P=0.001).
CONCLUSIONS: These results demonstrate the benefits of automated CVIP using smart pumps in limiting the frequency of haemodynamic incidents related to relays and in reducing the nursing workload.

Entities:  

Keywords:  care workload; critical care nursing; shock; smart pumps; vasoactive drugs

Mesh:

Substances:

Year:  2013        PMID: 23761221     DOI: 10.1093/bja/aet199

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  9 in total

1.  [Not Available].

Authors:  Emmanuelle Delage; Julien Tourel; Brigitte Martin; Aurélie Guérin; Ahmed Moussa; Annie Lacroix; Denis Lebel; Jean-François Bussières
Journal:  Can J Hosp Pharm       Date:  2015 Sep-Oct

2.  Performance of the BBraun perfusor space syringe driver under hyperbaric conditions.

Authors:  Lachlan Frawley; Bridget Devaney; Theo Tsouras; Geoff Frawley
Journal:  Diving Hyperb Med       Date:  2017-03       Impact factor: 0.887

3.  Accidents and Incidents Related to Intravenous Drug Administration: A Pre-Post Study Following Implementation of Smart Pumps in a Teaching Hospital.

Authors:  Aurélie Guérin; Julien Tourel; Emmanuelle Delage; Stéphanie Duval; Marie-Johanne David; Denis Lebel; Jean-François Bussières
Journal:  Drug Saf       Date:  2015-08       Impact factor: 5.606

4.  Long-term outcome of critically ill adult patients with acute epiglottitis.

Authors:  Tomasz Chroboczek; Martin Cour; Romain Hernu; Thomas Baudry; Julien Bohé; Vincent Piriou; Bernard Allaouchiche; François Disant; Laurent Argaud
Journal:  PLoS One       Date:  2015-05-06       Impact factor: 3.240

5.  Predictors of haemodynamic instability during the changeover of norepinephrine infusion pumps.

Authors:  Martin Cour; Thomas Bénet; Romain Hernu; Marie Simon; Thomas Baudry; Philippe Vanhems; Laurent Argaud
Journal:  Ann Intensive Care       Date:  2016-04-22       Impact factor: 6.925

6.  Improved usability of a multi-infusion setup using a centralized control interface: A task-based usability test.

Authors:  Frank Doesburg; Fokie Cnossen; Willem Dieperink; Wouter Bult; Anne Marie de Smet; Daan J Touw; Maarten W Nijsten
Journal:  PLoS One       Date:  2017-08-11       Impact factor: 3.240

7.  Comparison of Operator Workloads Associated with the Single-unit Anyfusion® Pump and the Changeover from a Syringe Pump to an Infusion Pump.

Authors:  Jeong Jin Min; Duk Kyung Kim; Kwan Young Hong; Ji Won Choi; Ka Young Choi
Journal:  J Korean Med Sci       Date:  2019-12-23       Impact factor: 2.153

8.  Systemic Defenses to Prevent Intravenous Medication Errors in Hospitals: A Systematic Review.

Authors:  Sini Karoliina Kuitunen; Ilona Niittynen; Marja Airaksinen; Anna-Riia Holmström
Journal:  J Patient Saf       Date:  2021-12-01       Impact factor: 2.243

9.  Automatic versus manual changeovers of norepinephrine infusion pumps in critically ill adults: a prospective controlled study.

Authors:  Emilie Greau; Jean-Baptiste Lascarrou; Aurélie Le Thuaut; Nathalie Maquigneau; Yolaine Alcourt; Anne Coutolleau; Cécile Rousseau; Vanessa Erragne; Jean Reignier
Journal:  Ann Intensive Care       Date:  2015-11-14       Impact factor: 6.925

  9 in total