Literature DB >> 237610

A hazard associated with improper connection of the Bain breathing circuit.

J G Paterson, V Vanhooydonk.   

Abstract

A case of cardiac arrest is presented which was caused by improper connection of a modified Mapleson D circuit (Bain breathing circuit). Excessive enternal deadspace was created by interchanging the gas inflow line and the attachment for an airway pressure manometer. This resulted in a marked respiratory acidosis, clinically undetected until the concomitant hypoxia produced a severe cardiac depression and arrest. The sequence of events was reproduced in a dog under comparable anaesthetic conditions. In order to avoid this error it is recommended to permanently fuse the connecting piece placed in the circuit with the tubing leading to the airway pressure manometer of the respirator.

Entities:  

Mesh:

Substances:

Year:  1975        PMID: 237610     DOI: 10.1007/bf03004847

Source DB:  PubMed          Journal:  Can Anaesth Soc J        ISSN: 0008-2856


  2 in total

1.  A hazard connected with re-use of the Bain's circuit: A case report.

Authors:  R Hannallah; J K Rosales
Journal:  Can Anaesth Soc J       Date:  1974-09

2.  A streamlined anaesthetic system.

Authors:  J A Bain; W E Spoerel
Journal:  Can Anaesth Soc J       Date:  1972-07
  2 in total
  7 in total

1.  Letter: The performance of a modified Mapleson D system.

Authors:  C M Conway
Journal:  Can Anaesth Soc J       Date:  1975-09

2.  Modifying nitrous oxide analgesia circuits to reduce pollution in the dental operatory.

Authors:  D Adams; G D Allen; J Scaramella
Journal:  Anesth Prog       Date:  1976 Nov-Dec

3.  Single lever Humphrey A.D.E. lowflow universal anaesthetic breathing system. Part II: Comparison with Bain system in anaesthesized adults during controlled ventilation.

Authors:  D Humphrey; J G Brock-Utne; J W Downing
Journal:  Can Anaesth Soc J       Date:  1986-11

4.  Single lever Humphrey A.D.E. low flow universal anaesthetic breathing system.

Authors:  J A Bain
Journal:  Can J Anaesth       Date:  1987-05       Impact factor: 5.063

5.  Carbon dioxide absorption system with a simple modification of the Bain circuit.

Authors:  K H Rah; N R Lee; M L McDougle; R L Campbell
Journal:  Anesth Prog       Date:  1982 Sep-Oct

6.  The use of the Bain circuit in spontaneously breathing paediatric patients.

Authors:  M G Soliman; R Laberge
Journal:  Can Anaesth Soc J       Date:  1978-07

7.  Newer design, newer problems: Unusual complication with Limb-O anaesthesia circuit.

Authors:  Chitra Rajeswari Thangaswamy; Patel Roushan; Lal Pooja; Guru Krishnakumar; Lenin Babu Elakkumanan
Journal:  Indian J Anaesth       Date:  2017-10
  7 in total