Literature DB >> 23760745

Incidence of adverse wear reactions in hip resurfacing arthroplasty: a single surgeon series of 2,600 cases.

Thomas P Gross1, Fei Liu.   

Abstract

A single surgeon performed 2,559 metal-on-metal hip resurfacing arthroplasties in 
2,109 patients. The Corin Cormet 2000 (393 cases) and Biomet Recap implants (2,166 cases) were used in our series. In this study, the adverse wear failure (AWF) rate was 0.27%. At 10 years postoperatively, our Kaplan-Meier cumulative revision rate for AWF was 1% for all patients, 0.2% for men, 2.6% for women, and 9% for patients with a diagnosis of dysplasia. All AWF failures had component sizes ≤48 mm. All had metal ion levels above 15 ug/ml. All had acetabular inclination angles (AIA) ≥50° on standing pelvis radiographs. All had severe metallosis found at the time of revision. Six of the seven AWF cases were in women. There were no failures from pseudotumours without AWF (metallosis) in this series.

Entities:  

Mesh:

Year:  2013        PMID: 23760745     DOI: 10.5301/hipint.5000030

Source DB:  PubMed          Journal:  Hip Int        ISSN: 1120-7000            Impact factor:   2.135


  5 in total

1.  Reply to letter to the editor: do complication rates differ by gender after metal-on-metal hip resurfacing arthroplasty? A systematic review.

Authors:  Bryan D Haughom; Brandon J Erickson; Michael D Hellman; Joshua J Jacobs
Journal:  Clin Orthop Relat Res       Date:  2015-09-23       Impact factor: 4.176

Review 2.  Do Complication Rates Differ by Gender After Metal-on-metal Hip Resurfacing Arthroplasty? A Systematic Review.

Authors:  Bryan D Haughom; Brandon J Erickson; Michael D Hellman; Joshua J Jacobs
Journal:  Clin Orthop Relat Res       Date:  2015-08       Impact factor: 4.176

3.  Surface Characterization of Retrieved Metal-on-Metal Total Hip Implants from Patients with Adverse Reaction to Metal Debris.

Authors:  Maria Burbano; Robert Russell; Michael Huo; Robert Welch; Diana Roy; Danieli C Rodrigues
Journal:  Materials (Basel)       Date:  2014-03-04       Impact factor: 3.623

Review 4.  Prevalence of Failure due to Adverse Reaction to Metal Debris in Modern, Medium and Large Diameter Metal-on-Metal Hip Replacements--The Effect of Novel Screening Methods: Systematic Review and Metaregression Analysis.

Authors:  Aleksi Reito; Olli Lainiala; Petra Elo; Antti Eskelinen
Journal:  PLoS One       Date:  2016-03-01       Impact factor: 3.240

5.  Poor 10-year survivorship of hip resurfacing arthroplasty.

Authors:  Matti Seppänen; Mikko Karvonen; Petri Virolainen; Ville Remes; Pekka Pulkkinen; Antti Eskelinen; Antti Liukas; Keijo T Mäkelä
Journal:  Acta Orthop       Date:  2016-10-19       Impact factor: 3.717

  5 in total