Fast methodology for the reliable determination of nonylphenol in water samples by minimal labeling isotope dilution mass spectrometry.

Neus Fabregat-Cabello1, Ángel Castillo, Juan V Sancho, Florenci V González, Antoni Francesc Roig-Navarro.   

Abstract

In this work we have developed and validated an accurate and fast methodology for the determination of 4-nonylphenol (technical mixture) in complex matrix water samples by UHPLC-ESI-MS/MS. The procedure is based on isotope dilution mass spectrometry (IDMS) in combination with isotope pattern deconvolution (IPD), which provides the concentration of the analyte directly from the spiked sample without requiring any methodological calibration graph. To avoid any possible isotopic effect during the analytical procedure the in-house synthesized (13)C1-4-(3,6-dimethyl-3-heptyl)phenol was used as labeled compound. This proposed surrogate was able to compensate the matrix effect even from wastewater samples. A SPE pre-concentration step together with exhaustive efforts to avoid contamination were included to reach the signal-to-noise ratio necessary to detect the endogenous concentrations present in environmental samples. Calculations were performed acquiring only three transitions, achieving limits of detection lower than 100ng/g for all water matrix assayed. Recoveries within 83-108% and coefficients of variation ranging from 1.5% to 9% were obtained. On the contrary a considerable overestimation was obtained with the most usual classical calibration procedure using 4-n-nonylphenol as internal standard, demonstrating the suitability of the minimal labeling approach.
Copyright © 2013 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Isotope dilution mass spectrometry; Isotope pattern deconvolution; LC–MS/MS; Minimal labeling; Nonylphenol; Wastewater analysis

Mesh:

Substances:

Year:  2013        PMID: 23746647     DOI: 10.1016/j.chroma.2013.05.033

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  1 in total

1.  Application of screening experimental designs to assess chromatographic isotope effect upon isotope-coded derivatization for quantitative liquid chromatography-mass spectrometry.

Authors:  Szabolcs Szarka; Katalin Prokai-Tatrai; Laszlo Prokai
Journal:  Anal Chem       Date:  2014-06-26       Impact factor: 6.986

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.