Literature DB >> 23742557

Sensorless enhancement of an atomic force microscope micro-cantilever quality factor using piezoelectric shunt control.

M Fairbairn1, S O R Moheimani.   

Abstract

The image quality and resolution of the Atomic Force Microscope (AFM) operating in tapping mode is dependent on the quality (Q) factor of the sensing micro-cantilever. Increasing the cantilever Q factor improves image resolution and reduces the risk of sample and cantilever damage. Active piezoelectric shunt control is introduced in this work as a new technique for modifying the Q factor of a piezoelectric self-actuating AFM micro-cantilever. An active impedance is placed in series with the tip oscillation voltage source to modify the mechanical dynamics of the cantilever. The benefit of using this control technique is that it removes the optical displacement sensor from the Q control feedback loop to reduce measurement noise in the loop and allows for a reduction in instrument size.

Entities:  

Year:  2013        PMID: 23742557     DOI: 10.1063/1.4805108

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523


  1 in total

1.  Squeeze Film Damping Effect on Different Microcantilever Probes in Tapping Mode Atomic Force Microscope.

Authors:  Yan Sun; Jing Liu; Kejian Wang; Zheng Wei
Journal:  Scanning       Date:  2020-11-13       Impact factor: 1.932

  1 in total