Optical design of the short pulse x-ray imaging and microscopy time-angle correlated diffraction beamline at the Advanced Photon Source.

R Reininger1, E M Dufresne, M Borland, M A Beno, L Young, K-J Kim, P G Evans.   

Abstract

The short pulse x-ray imaging and microscopy beamline is one of the two x-ray beamlines that will take full advantage of the short pulse x-ray source in the Advanced Photon Source (APS) upgrade. A horizontally diffracting double crystal monochromator which includes a sagittally focusing second crystal will collect most of the photons generated when the chirped electron beam traverses the undulator. A Kirkpatrick-Baez mirror system after the monochromator will deliver to the sample a beam which has an approximately linear correlation between time and vertical beam angle. The correlation at the sample position has a slope of 0.052 ps/μrad extending over an angular range of 800 μrad for a cavity deflection voltage of 2 MV. The expected time resolution of the whole system is 2.6 ps. The total flux expected at the sample position at 10 keV with a 0.9 eV energy resolution is 5.7 × 10(12) photons/s at a spot having horizontal and vertical full width at half maximum of 33 μm horizontal by 14 μm vertical. This new beamline will enable novel time-dispersed diffraction experiments on small samples using the full repetition rate of the APS.

Entities:  

Year:  2013        PMID: 23742528     DOI: 10.1063/1.4804197

Source DB:  PubMed          Journal:  Rev Sci Instrum        ISSN: 0034-6748            Impact factor:   1.523


  1 in total

1.  Optical design of the Short Pulse Soft X-ray Spectroscopy beamline at the Advanced Photon Source.

Authors:  R Reininger; D J Keavney; M Borland; L Young
Journal:  J Synchrotron Radiat       Date:  2013-05-30       Impact factor: 2.616

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.