Electrical features of eighteen automated external defibrillators: a systematic evaluation.

Fulvio Kette1, Aldo Locatelli, Marcella Bozzola, Alberto Zoli, Yongqin Li, Marco Salmoiraghi, Giuseppe Ristagno, Aida Andreassi.   

Abstract

AIM: Assessment and comparison of the electrical parameters (energy, current, first and second phase waveform duration) among eighteen AEDs.
METHOD: Engineering bench tests for a descriptive systematic evaluation in commercially available AEDs. AEDs were tested through an ECG simulator, an impedance simulator, an oscilloscope and a measuring device detecting energy delivered, peak and average current, and duration of first and second phase of the biphasic waveforms. All tests were performed at the engineering facility of the Lombardia Regional Emergency Service (AREU).
RESULTS: Large variations in the energy delivered at the first shock were observed. The trend of current highlighted a progressive decline concurrent with the increases of impedance. First and second phase duration varied substantially among the AEDs using the exponential biphasic waveform, unlike rectilinear waveform AEDs in which phase duration remained relatively constant.
CONCLUSIONS: There is a large variability in the electrical features of the AEDs tested. Energy is likely not to be the best indicator for strength dose selection. Current and shock duration should be both considered when approaching the technical features of AEDs. These findings may prompt further investigations to define the optimal current and duration of the shock waves to increase the success rate in the clinical setting.
Copyright © 2013 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Keywords:  Automated external defibrillators; Current; Energy; Ventricular fibrillation; Waveform

Mesh:

Year:  2013        PMID: 23735652     DOI: 10.1016/j.resuscitation.2013.05.017

Source DB:  PubMed          Journal:  Resuscitation        ISSN: 0300-9572            Impact factor:   5.262


  2 in total

1.  A framework of current based defibrillation improves defibrillation efficacy of biphasic truncated exponential waveform in rabbits.

Authors:  Weiming Li; Jingru Li; Liang Wei; Jianjie Wang; Li Peng; Juan Wang; Changlin Yin; Yongqin Li
Journal:  Sci Rep       Date:  2021-01-15       Impact factor: 4.379

2.  First-time evaluation of ascending compared to rectangular transthoracic defibrillation waveforms in modelled out-of-hospital cardiac arrest.

Authors:  Tobias Neumann; Simon-Richard Finke; Maja Henninger; Sebastian Lemke; Ben Hoepfner; Daniel Steven; Alexandra C Maul; Daniel C Schroeder; Thorsten Annecke
Journal:  Resusc Plus       Date:  2020-06-01
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.