BACKGROUND: Accurate patient to image registration is the core for successful image-guided neurosurgery. While skin adhesive markers (SMs) are widely used in point-matching registration, a proper implementation of anatomical landmarks (ALs) may overcome the inconvenience brought by the use of SMs.
METHODS: Using nine ALs, a set of three configurations of different combinations of them is proposed. These configurations are defined according to the required positioning of the patient's head during surgery and the resulting distribution of the expected target registration error (TRE). These configurations were first evaluated by simulation experiment using the data of 20 patients from two hospitals, and then testing the applicability of them in eight real clinical surgeries of neuronavigation.
RESULTS: The results of the simulation experiment showed that, by incorporating a fiducial registration error (FRE) of 3.5 mm measured in the clinical setting, the expected TRE in the whole skull was less than 2.5 mm, and the expected TRE in the whole brain was less than 1.75 mm when using all the nine ALs. A small TRE could also be achieved in the corresponding surgical field by using the other three configurations with less ALs. In the clinical experiment, the FLE ranges in the image and the patient space were 1.4-3.6 mm and 1.6-5.5 mm, respectively. The measured TRE and FRE were 3.1 ± 0.75 mm and 3.5 ± 0.17 mm, respectively.
CONCLUSIONS: The AL configurations proposed in this investigation provide sufficient registration accuracy and can help to avoid the disadvantages of SMs if used clinically.
Copyright © 2013 John Wiley & Sons, Ltd.

Entities:  

Keywords:  anatomical landmarks; image-guided neurosurgery; point-matching registration; target registration error

Mesh:

Year:  2013        PMID: 23733606     DOI: 10.1002/rcs.1509

Source DB:  PubMed          Journal:  Int J Med Robot        ISSN: 1478-5951            Impact factor:   2.547


  2 in total

1.  Strategy for accurate liver intervention by an optical tracking system.

Authors:  Qinyong Lin; Rongqian Yang; Ken Cai; Peifeng Guan; Weihu Xiao; Xiaoming Wu
Journal:  Biomed Opt Express       Date:  2015-08-07       Impact factor: 3.732

2.  How precise are oral splints for frameless stereotaxy in guided ear, nose, throat, and maxillofacial surgery: a cadaver study.

Authors:  Manfred Nilius; Minou Hélène Nilius
Journal:  Eur Radiol Exp       Date:  2021-06-30
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.